FDA Adverse Event Malfunction Summary report: N

NAVLOCK

MDR report key: 24533467 · Received March 6, 2026

Report

Report Number
1723170-2026-00369
Event Type
Malfunction
Date Received
March 6, 2026
Date of Event
April 2, 2024
Report Date
March 6, 2026
Manufacturer
MEDTRONIC NAVIGATION, INC.
Product Code
OLO
PMA / PMN Number
K124004
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

H3, H6; A MEDTRONIC REPRESENTATIVE WENT TO THE SITE TO TEST THE EQUIPMENT. THE SHARP AWL WAS REPLACED AND THE SYSTEM THEN PASSED TES TING. CODES B01, C13, AND D02 ARE APPLICABLE TO THIS ANALYSIS. THE SHARP AWL, LOT NUMBER: 221218, WAS RETURNED TO THE MANUFACTURER FOR ANALYSIS. THE TIP OF THE RETURNED SHARP AWL HAS BEEN BLUNTED. CODES B01, C13, AND D02 ARE APPLICABLE TO THIS ANALYSIS. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Description of Event or Problem · 0

MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION SYSTEM BEING USED OUTSIDE OF A PROCEDURE. IT WAS REPORTED THAT THERE WAS DAMAGE TO THE HEAD OF AN INSTRUMENT. THERE WAS NO PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
592377 NAVLOCK ORTHOPEDIC STEREOTAXIC INSTRUMENT OLO MEDTRONIC NAVIGATION, INC. 9734678 00763000054182

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown