FDA Adverse Event
Malfunction
Summary report: N
I-STAT TROPONIN (CTNI) CARTRIDGE
MDR report key: 2453250
·
Received January 23, 2012
Report
- Report Number
- 2245578-2012-00034
- Event Type
- Malfunction
- Date Received
- January 23, 2012
- Date of Event
- December 28, 2011
- Report Date
- January 4, 2012
- Manufacturer
- ABBOTT POINT OF CARE INC.
- Product Code
- MMI
- PMA / PMN Number
- K031739
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WV, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
(B)(4).
Description of Event or Problem · 1
ON (B)(6) 2012, ABBOTT POINT OF CARE WAS CONTACTED BY A CUSTOMER REGARDING I-STAT TROPONIN CARTRIDGES THAT YIELDED DISCREPANT RESULT OF 0.23 ON A PT. SAMPLE COLLECTION: (B)(6) 2011, 15:45 SAMPLE A COLLECTION; (B)(6) 2011 17:23 SAMPLE B COLLECTION. I-STAT, (B)(6) 2011 16:14 0.23 NG/ML (SAMPLE A); (B)(6) 2011 16:43 0.02 NG/ML (SAMPLE A); (B)(6) 2011 17:26 0.01 NG/ML (SAMPLE B). CENTAUR (LAB), (B)(6) 2011 20:34 <0.006 NG/ML; (B)(6) 2011 04:20 <0.006 NG/ML. BASED ON THE INFO AVAILABLE, THERE WERE NO INJURIES ASSOCIATED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | I-STAT TROPONIN (CTNI) CARTRIDGE | CTNI CARTRIDGE | MMI | ABBOTT POINT OF CARE INC. | NA | U11270 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 52 YR |