FDA Adverse Event Malfunction Summary report: N

DAVINCI XI

MDR report key: 24526056 · Received March 5, 2026

Report

Report Number
2955842-2026-15178
Event Type
Malfunction
Date Received
March 5, 2026
Date of Event
February 9, 2026
Report Date
March 5, 2026
Manufacturer
INTUITIVE SURGICAL, INC
Product Code
NAY
UDI-DI
00886874110898
PMA / PMN Number
K131861
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

AN INVESTIGATION WAS COMPLETED TO DETERMINE THE CAUSE OF THIS REPORTED EVENT. AN INTUITIVE SURGICAL, INC. (ISI) FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO THE CUSTOMER SITE TO FURTHER INVESTIGATE THE REPORTED EVENT. CORE POWER TRAY ASSEMBLY HAS BEEN REPLACED TO CORRECT THE REPORTED ERROR. THE SYSTEM WAS TESTED AND VERIFIED AS READY FOR USE. ISI DID NOT RECEIVE A DA VINCI PRODUCT INVOLVED WITH THIS COMPLAINT TO PERFORM FAILURE ANALYSIS. THE COMPLAINT WAS CONFIRMED BASED ON THE FIELD EVALUATION, WHICH INDICATES THAT THE DEVICE MAY HAVE CONTRIBUTED TO THE CUSTOMER REPORTED ISSUE.

Description of Event or Problem · 0

IT WAS REPORTED THAT PRIOR TO THE START OF A DA VINCI-ASSISTED NEPHRECTOMY SURGICAL PROCEDURE, USER INFORMED THE TECHNICAL SUPPORT ENGINEER (TSE) THAT THEY ENCOUNTERED A REPEATED NON-RECOVERABLE ERROR 8 . THE USER ATTEMPTED MULTIPLE SYSTEM RESTARTS, BUT THE ISSUE PERSISTED. THE TSE CONDUCTED A VIDEO CALL TO GUIDE THE USER THROUGH SHUTTING DOWN THE SYSTEM, SWITCHING OFF THE CIRCUIT BREAKER TO THE VISION SIDE CART, ENSURING INTERNAL MAINS CABLES WERE SECURE, CYCLING THE CORE POWER SWITCH, AND RETURNING ALL SWITCHES TO THE ON POSITION. DESPITE THESE EFFORTS, THE ERROR CONDITION REMAINED UNCHANGED. THE PROCEDURE WAS ABORTED WITH NO PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
586592 DAVINCI XI VISION SIDE SYSTEM NAY INTUITIVE SURGICAL, INC 381121-41 N/A 00886874110898

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown DA VINCI INSTRUMENTS AND ACCESSORIES