FDA Adverse Event Malfunction Summary report: N

DAVINCI XI

MDR report key: 24525736 · Received March 5, 2026

Report

Report Number
2955842-2026-15167
Event Type
Malfunction
Date Received
March 5, 2026
Date of Event
February 9, 2026
Report Date
April 7, 2026
Manufacturer
INTUITIVE SURGICAL, INC
Product Code
NAY
UDI-DI
00886874110744
PMA / PMN Number
K131861
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

INTUITIVE SURGICAL, INC. (ISI) DID RECEIVE THE HIGH RESOLUTION STEREO VIEWER (HRSV) TO PERFORM FAILURE ANALYSIS AND THE REPORTED FAILURE WAS REPLICATED DURING IN-HOUSE TESTING AND CONFIRMED VIA SYSTEM LOGS. UPON VISUAL INSPECTION, NO ISSUE WAS FOUND THAT WOULD RELATE TO THE COMPLAINT. THE HRSV WAS THEN INSTALLED ONTO THE GOLDEN SYSTEM. UPON POWERING UP THE SYSTEM, THERE WAS NO IMAGE ON THE HRSV, IT WAS COMPLETELY BLACK. THE PROBABLE ROOT CAUSE IS ATTRIBUTED TO DISPLAY DEFECT PERTAINING TO THE HRSV MONITOR.

Additional Manufacturer Narrative · 0

AN INTUITIVE SURGICAL, INC. (ISI) FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO THE CUSTOMER SITE TO FURTHER INVESTIGATE THE REPORTED COMPLAINT. THE FSE WAS ABLE TO REPRODUCE THE ISSUE AND REPLACED THE HIGH RESOLUTION STEREO VIEWER (HRSV). THE SYSTEM WAS TESTED AND VERIFIED AS READY FOR USE. AS OF THE DATE OF THIS REPORT, THE HRSV HAS NOT YET BEEN RECEIVED BY ISI FOR EVALUATION.

Description of Event or Problem · 0

IT WAS REPORTED THAT PRIOR TO STARTING A DA VINCI-ASSISTED SURGICAL PROCEDURE, THERE WAS NO IMAGE DISPLAYED FOR THE LEFT HIGH RESOLUTION STEREO VIEWER (HRSV), AND THE DA VINCI LOGO DID NOT APPEAR. A HARD POWER CYCLE DID NOT RESOLVE THE ISSUE. THE PROCEDURE WAS ABORTED, AND NO PATIENT INVOLVEMENT WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
446214 DAVINCI XI SURGEON SIDE CONSOLE, SMART PEDALS NAY INTUITIVE SURGICAL, INC 380677-10 N/A 00886874110744

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown DA VINCI INSTRUMENTS AND ACCESSORIES.