DAVINCI XI
Report
- Report Number
- 2955842-2026-15167
- Event Type
- Malfunction
- Date Received
- March 5, 2026
- Date of Event
- February 9, 2026
- Report Date
- April 7, 2026
- Manufacturer
- INTUITIVE SURGICAL, INC
- Product Code
- NAY
- UDI-DI
- 00886874110744
- PMA / PMN Number
- K131861
- Removal / Correction Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- 003
Narratives
INTUITIVE SURGICAL, INC. (ISI) DID RECEIVE THE HIGH RESOLUTION STEREO VIEWER (HRSV) TO PERFORM FAILURE ANALYSIS AND THE REPORTED FAILURE WAS REPLICATED DURING IN-HOUSE TESTING AND CONFIRMED VIA SYSTEM LOGS. UPON VISUAL INSPECTION, NO ISSUE WAS FOUND THAT WOULD RELATE TO THE COMPLAINT. THE HRSV WAS THEN INSTALLED ONTO THE GOLDEN SYSTEM. UPON POWERING UP THE SYSTEM, THERE WAS NO IMAGE ON THE HRSV, IT WAS COMPLETELY BLACK. THE PROBABLE ROOT CAUSE IS ATTRIBUTED TO DISPLAY DEFECT PERTAINING TO THE HRSV MONITOR.
AN INTUITIVE SURGICAL, INC. (ISI) FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO THE CUSTOMER SITE TO FURTHER INVESTIGATE THE REPORTED COMPLAINT. THE FSE WAS ABLE TO REPRODUCE THE ISSUE AND REPLACED THE HIGH RESOLUTION STEREO VIEWER (HRSV). THE SYSTEM WAS TESTED AND VERIFIED AS READY FOR USE. AS OF THE DATE OF THIS REPORT, THE HRSV HAS NOT YET BEEN RECEIVED BY ISI FOR EVALUATION.
IT WAS REPORTED THAT PRIOR TO STARTING A DA VINCI-ASSISTED SURGICAL PROCEDURE, THERE WAS NO IMAGE DISPLAYED FOR THE LEFT HIGH RESOLUTION STEREO VIEWER (HRSV), AND THE DA VINCI LOGO DID NOT APPEAR. A HARD POWER CYCLE DID NOT RESOLVE THE ISSUE. THE PROCEDURE WAS ABORTED, AND NO PATIENT INVOLVEMENT WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 446214 | DAVINCI XI | SURGEON SIDE CONSOLE, SMART PEDALS | NAY | INTUITIVE SURGICAL, INC | 380677-10 | N/A | 00886874110744 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown | DA VINCI INSTRUMENTS AND ACCESSORIES. |