FDA Adverse Event Injury Summary report: N

MENTOR SMOOTH ROUND MODERATE PROFILE SALINE BREAST PROSTHESIS

MDR report key: 24524390 · Received March 5, 2026

Report

Report Number
1645337-2026-02468
Event Type
Injury
Date Received
March 5, 2026
Date of Event
August 13, 2025
Manufacturer
MENTOR TEXAS
Product Code
FWM
UDI-DI
00081317001232
PMA / PMN Number
P990075
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

ON MARCH 06, 2026, THE MENTOR FAILURE ANALYSIS LAB HAS RECEIVED THE DEVICE FOR EVALUATION. THE ANALYSIS HAS BEGUN BUT IS NOT COMPLETE AT THIS TIME. WHEN THE INVESTIGATIONAL ANALYSIS HAS BEEN COMPLETED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THE IMPACTED DEVICE WAS DETERMINED TO BE A 275CC MENTOR SMOOTH ROUND MODERATE PROFILE SALINE BREAST PROSTHESIS WITH LOT NUMBER 5925416. A MANUFACTURING RECORD EVALUATION (MRE) WAS PERFORMED, AND NO ANOMALIES WERE FOUND RELATED TO THIS COMPLAINT. IN ADDITION, THE MRE VERIFIES THAT THE DEVICE WAS MANUFACTURED IN ACCORDANCE WITH DOCUMENTED SPECIFICATION AND PROCEDURES. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Additional Manufacturer Narrative · 0

D4: UDI: AS THE SERIAL NUMBER FOR THE DEVICE INVOLVED IN THE EVENT WAS NOT PROVIDED, THE FULL UDI IS CURRENTLY NOT AVAILABLE. ON MARCH 25, 2026, THE EVALUATION FOR THE DEVICE WAS COMPLETED. DEVICE EVALUATION SUMMARY: VISUAL ANALYSIS OF THE RETURNED DEVICE REVEALED THAT THE BREAST IMPLANT HAD A CREASE/FOLD FROM THE POSTERIOR VIEW EXTENDING TO THE ANTERIOR VIEW. LEAK TESTING WAS PERFORMED, ACCORDING TO MENTOR PROCEDURE, AND IT IDENTIFIED A TEAR WITHIN THE CREASE/FOLD ON THE POSTERIOR VIEW, MEASURING APPROXIMATELY 0.3 CM. NO MORE LEAK SITES WERE OBSERVED. THE EVALUATION DETERMINED THAT THE POSSIBLE CAUSE OF THE RUPTURE IS CONSISTENT WITH NORMAL WEAR. DEFLATION CAN OCCUR AT ANY TIME AFTER IMPLANTATION, BUT IT IS MORE LIKELY TO OCCUR THE LONGER THE IMPLANT IS IMPLANTED. CREATING WRINKLES OR FOLDS SHOULD BE AVOIDED DURING IMPLANTATION OR OTHER PROCEDURES AS IT CAN RESULT IN A DEFLATION IN A LATER TIME. BREAST IMPLANTS MAY ALSO SIMPLY WEAR OUT OVER TIME. AS PART OF MENTOR'S QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED ACCORDING TO APPROVED SPECIFICATIONS. BASED ON THE RESULTS OF THIS INVESTIGATION, IT WAS DETERMINED THAT THE PRODUCT MET THE MANUFACTURING RELEASE CRITERIA, AND THAT NORMAL WEAR MAY HAVE CAUSED THE OBSERVED DEFECT. NO CORRECTIVE AND PREVENTIVE ACTION (CAPA) IS REQUIRED NOW. ADDITIONAL COMPLAINT INFORMATION MONITORING FOR POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AS PART OF POST MARKET SURVEILLANCE. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Additional Manufacturer Narrative · 0

SINCE THE DEVICE HAS NOT BEEN RETURNED FOR ANALYSIS, NO PRODUCT FAILURE ANALYSIS CAN BE CONDUCTED, AND NO DETERMINATION OF POSSIBLE CONTRIBUTING FACTORS CAN BE MADE. AS SUCH, THE INVESTIGATION WILL BE CLOSED. IF THE COMPLAINT DEVICE IS RECEIVED IN THE FUTURE, THE INVESTIGATION WILL BE REOPENED AND CONDUCTED AS APPROPRIATE. SINCE NO LOT NUMBER WAS PROVIDED, NO MANUFACTURING RECORD EVALUATION COULD BE PERFORMED. REASON FOR DEVICE EXPLANT AND/OR REOPERATION: DEFLATION. D4: UDI: AS THE LOT/SERIAL NUMBER FOR THE DEVICE INVOLVED IN THE EVENT WAS NOT PROVIDED, THE FULL UDI IS CURRENTLY NOT AVAILABLE. MENTOR IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH MENTOR HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, MENTOR, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, MENTOR, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF CERTAIN INFORMATION IS UNKNOWN, NOT AVAILABLE OR DOES NOT APPLY, THE SECTION/FIELD OF THE FORM IS LEFT BLANK. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Description of Event or Problem · 0

IT WAS REPORTED THAT A FEMALE PATIENT UNDERWENT A BREAST AUGMENTATION REVISION WITH A 275CC MENTOR SMOOTH ROUND MODERATE PROFILE SALINE BREAST PROSTHESIS AND EXPERIENCED A DEFLATION ON THE LEFT SIDE POST-OPERATIVELY. AS A RESULT, THE PATIENT UNDERWENT EXPLANTATION ON (B)(6) 2025.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
579449 MENTOR SMOOTH ROUND MODERATE PROFILE SALINE BREAST PROSTHESIS PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE FWM MENTOR TEXAS 5925416 00081317001232

Patients

Seq Age Sex Outcome Treatment
1 67 YR Female Required Intervention