FDA Adverse Event Malfunction Summary report: N

QUANTUM ROLLER PUMP - 6"

MDR report key: 24519752 · Received March 5, 2026

Report

Report Number
3006073153-2026-00047
Event Type
Malfunction
Date Received
March 5, 2026
Date of Event
March 8, 2023
Report Date
March 5, 2026
Manufacturer
SPECTRUM MEDICAL LTD
Product Code
DTQ
UDI-DI
05060434420787
PMA / PMN Number
K173834
Removal / Correction Number
Z-0224-2025
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION CONFIRM THE LEFT SIDE BOBBINS ARE STRIPPED AND WOULD NOT STAY LOCKED. THIS EVENT WAS ORIGINALLY REPORTED IN A SUMMARY REPORT 3006073153-2025-00059, WHICH WAS SUBMITTED IN RELATION TO THE AFFECTED DEVICES FROM THE VOLUNTARY RECALLED PERFORMED BY SPECTRUM MEDICAL LTD (RECALL NUMBER Z-0224-2025). FDA HAS SINCE REQUESTED THAT EACH EVENT FOR THE DEVICES RELATED TO THE VOLUNTARY RECALL BE INDIVIDUALLY SUBMITTED, RESULTING IN THE CREATION AND SUBMISSION OF THIS REPORT.

Description of Event or Problem · 0

USER REPORTED THAT WHEN BOBBIN LOCKED, RATCHETS WOULD NOT REMAIN LOCKED IN PLACE AND TUBING WAS NOT SECURED TIGHTLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
583691 QUANTUM ROLLER PUMP - 6" ROLLER PUMP DTQ SPECTRUM MEDICAL LTD QRP6 05060434420787

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown