FDA Adverse Event Malfunction Summary report: N

COBAS PRO ISE ANALYTICAL UNIT

MDR report key: 24518347 · Received March 5, 2026

Report

Report Number
1823260-2026-00775
Event Type
Malfunction
Date Received
March 5, 2026
Date of Event
February 11, 2026
Report Date
April 27, 2026
Manufacturer
ROCHE DIAGNOSTICS
Product Code
JJE
PMA / PMN Number
K191899
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE COBAS PRO ISE ANALYZER SERIAL NUMBER WAS (B)(6).

Additional Manufacturer Narrative · 0

SECTION D, DEVICE IDENTIFICATION, WAS UPDATED. RELEVANT FIELDS OF SECTIONS D AND G WERE UPDATED. THE NA ELECTRODE WAS EQL WITH AN EXPIRATION DATE OF 27-JUL-2026.

Description of Event or Problem · 0

THE INITIAL REPORTER COMPLAINED OF QC ISSUES ON A COBAS PRO ISE ANALYTICAL UNIT. THE CUSTOMER REPLACED THE ELECTRODES, PERFORMED A PROBE WASH, COMPLETED CONDITIONING, AND PERFORMED PRECISION TESTING, WITH ACCEPTABLE RESULTS. THE CUSTOMER PERFORMED A REVIEW OF PATIENT RESULTS THAT HAD BEEN REPORTED OUTSIDE OF THE LABORATORY AND PERFORMED REPEAT TESTING ON A DIFFERENT COBAS PRO ISE ANALYZER. DISCREPANT RESULTS WERE IDENTIFIED FOR 4 PATIENT SAMPLES TESTED FOR SODIUM ELECTRODE (NA) AND ISE INDIRECT NA-K-CL FOR GEN.2 (CL). THIS MEDWATCH WILL COVER NA. REFER TO MEDWATCH WITH A1 PATIENT IDENTIFIER PT-0084466 FOR INFORMATION ON THE CL RESULTS. PATIENT 1 INITIAL NA RESULT WAS 156 MMOL/L. THE REPEAT RESULT WAS 142 MMOL/L. THE INITIAL CL RESULT WAS 113 MMOL/L. THE REPEAT RESULT WAS 102 MMOL/L. PATIENT 2 INITIAL NA RESULT WAS 148 MMOL/L. THE REPEAT RESULT WAS 139 MMOL/L. PATIENT 3 INITIAL NA RESULT WAS 154 MMOL/L. THE REPEAT RESULT WAS 147 MMOL/L. PATIENT 4 INITIAL NA RESULT WAS 151 MMOL/L. THE REPEAT RESULT WAS 139 MMOL/L. THE REPEAT RESULTS WERE BELIEVED TO BE CORRECT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
626618 COBAS PRO ISE ANALYTICAL UNIT CLINICAL CHEMISTRY ANALYZER JJE ROCHE DIAGNOSTICS EQL

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown