FDA Adverse Event Injury Summary report: N

COVEREDGE? 32

MDR report key: 24517638 · Received March 5, 2026

Report

Report Number
3006630150-2026-01282
Event Type
Injury
Date Received
March 5, 2026
Date of Event
January 13, 2026
Report Date
March 30, 2026
Manufacturer
BOSTON SCIENTIFIC NEUROMODULATION CORPORATION
Product Code
LGW
UDI-DI
08714729832669
PMA / PMN Number
P030017
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
UT, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

BLOCK B3: EXACT DATE APPROXIMATED BASED ON THE DATE THE MANUFACTURER BECAME AWARE OF THE EVENT. INVESTIGATION RESULTS: THE DEVICE WAS NOT RETURNED FOR ANALYSIS; THEREFORE, A TECHNICAL ANALYSIS COULD NOT BE PERFORMED. DEVICE HISTORY RECORD: IT WAS CONFIRMED THAT THIS DEVICE MET MANUFACTURING SPECIFICATION PRIOR TO DISTRIBUTION AND THERE WERE NO MANUFACTURING DEVIATIONS WHICH COULD HAVE CONTRIBUTED TO THE REPORTED EVENT.

Additional Manufacturer Narrative · 0

INVESTIGATION RESULTS: THE DEVICE WAS NOT RETURNED FOR ANALYSIS; THEREFORE, A TECHNICAL ANALYSIS COULD NOT BE PERFORMED. DEVICE HISTORY RECORD: IT WAS CONFIRMED THAT THIS DEVICE MET MANUFACTURING SPECIFICATION PRIOR TO DISTRIBUTION AND THERE WERE NO MANUFACTURING DEVIATIONS WHICH COULD HAVE CONTRIBUTED TO THE REPORTED EVENT. LABELING REVIEW: REVIEW OF THE INSTRUCTIONS FOR USE (IFU) CONFIRMED THERE WAS RELEVANT CONTENT AND SUFFICIENT GUIDANCE WITH RESPECT TO THE CIRCUMSTANCES DESCRIBED WITHIN THIS COMPLAINT. NO UPDATES ARE REQUIRED TO THE IFU AS A RESULT OF THIS EVENT. INVESTIGATION CONCLUSION: THE DEVICE WAS NOT RETURNED FOR ANALYSIS AND THE COMPLAINT AS REPORTED COULD NOT BE CONFIRMED. RECORD REVIEW REVEALED NO ADDITIONAL INFORMATION RELATED TO THE COMPLAINT. THEREFORE, THIS INVESTIGATION IS UNABLE TO DETERMINE A PROBABLE CAUSE FOR THE COMPLAINT AND THE CONCLUSION CODE KNOWN INHERENT RISK OF DEVICE WILL BE USED. CORRECTION TO FIELD H11: ADDITIONAL SUSPECT MEDICAL DEVICE COMPONENT INVOLVED IN THE EVENT: MODEL NUMBER/CATALOG NUMBER: SC-1232. SERIAL NUMBER: (B)(6). BATCH/LOT NUMBER: 770855. MODEL/CATALOG DESCRIPTION: WAVEWRITER ALPHA 32 IPG KIT. UNIQUE IDENTIFIER (UDI): (B)(4).

Description of Event or Problem · 0

IT WAS REPORTED THAT THE PATIENT WAS EXPERIENCING INADEQUATE STIMULATION DUE TO SPINAL CORD STIMULATOR (SCS) PADDLE LEAD HAD PULLED OUT OF THE EPIDURAL SPACE WHICH WAS CONFIRMED THROUGH XRAY. THE PATIENT UNDERWENT A PROCEDURE WHEREIN THE LEAD WAS REPLACED AND THAT THE PATIENT WAS DOING WELL POSTOPERATIVELY. THE EXPLANTED LEAD WAS RETAINED BY THE MEDICAL FACILITY AND WILL NOT BE RETURNED.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE PATIENT WAS EXPERIENCING INADEQUATE STIMULATION DUE TO SPINAL CORD STIMULATOR (SCS) PADDLE LEAD HAD PULLED OUT OF THE EPIDURAL SPACE WHICH WAS CONFIRMED THROUGH XRAY. THE PATIENT UNDERWENT A PROCEDURE WHEREIN THE LEAD WAS REPLACED AND THAT THE PATIENT WAS DOING WELL POSTOPERATIVELY. THE EXPLANTED LEAD WAS RETAINED BY THE MEDICAL FACILITY AND WILL NOT BE RETURNED. ADDITIONAL INFORMATION WAS RECEIVED THAT THE IPG WAS REPLACED ALONG WITH THE PADDLE LEAD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
111377 COVEREDGE? 32 STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF LGW BOSTON SCIENTIFIC NEUROMODULATION CORPORATION SC-8336-50 7072284 08714729832669

Patients

Seq Age Sex Outcome Treatment
1