DAVINCI XI
Report
- Report Number
- 2955842-2026-15066
- Event Type
- Malfunction
- Date Received
- March 4, 2026
- Date of Event
- February 11, 2026
- Report Date
- April 10, 2026
- Manufacturer
- INTUITIVE SURGICAL, INC
- Product Code
- NAY
- UDI-DI
- 00886874110744
- PMA / PMN Number
- K131861
- Removal / Correction Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
AN INVESTIGATION WAS COMPLETED TO DETERMINE THE CAUSE OF THIS REPORTED EVENT. AN INTUITIVE SURGICAL, INC. (ISI) FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO THE CUSTOMER SITE TO FURTHER INVESTIGATE THE REPORTED EVENT. FSE IDENTIFIED THAT PROBLEM WAS WITH THE PERSONALITY MODULE SURGEON CONSOLE (PMSC) BOARD IN SURGEON SID CONSOLE (SSC1) AND REPLACED THE PMSC. THE SYSTEM WAS TESTED AND VERIFIED AS READY FOR USE. THE COMPLAINT WAS CONFIRMED BASED ON FIELD EVALUATION. ISI HAS NOT RECEIVED THE PMSC FOR EVALUATION. A FOLLOW-UP MDR WILL BE SUBMITTED, IF ADDITIONAL INFORMATION IS OBTAINED.
INTUITIVE SURGICAL, INC. (ISI) DID RECEIVE A DA VINCI PRODUCT INVOLVED WITH THIS COMPLAINT TO PERFORM FAILURE ANALYSIS. THE PERSONALITY MODULE SURGEON CONSOLE (PMSC) WAS ANALYZED. ERROR 607 WAS FOUND IN ARTEMIS CONFIRMING THE REPORTED PROBLEM FROM THE FIELD. A VISUAL INSPECTION FOUND BOTH VIDEO INPUT LEAF (VIL) BOARDS AND SURGEON CONSOLE LEAF (SCL) BOARDS DAMAGED DUE TO WEAR AND TEAR. THE PMSC WAS INSTALLED ONTO AN IN-HOUSE SYSTEM WHERE BOTH THE VIL BOARDS AND SCL BOARDS WERE NOT PRESENT ON LAPTOP APP. ALSO, UPON POWERING UP, ERROR 48100 OCCURRED INDICATING THAT THERE WAS AN AUDIO ISSUE. THE COMPLAINT WAS CONFIRMED BASED ON FAILURE ANALYSIS. THE PROBABLE ROOT CAUSE IS ATTRIBUTED TO THE VIL BOARDS AND SCL BOARDS. THIS ISSUE CAN BE RESOLVED BY HAVING THE FSE REPLACE THE PSMC.
IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED RADICAL CYSTECTOMY W/ OTHER URINARY DIVERSION SURGICAL PROCEDURE, USER INFORMED THE TECHNICAL SUPPORT ENGINEER (TSE) THAT A REPEATED RECOVERABLE FAULT 607 OCCURRED. THE TSE REVIEWED THE LOGS AND INSTRUCTED THE USER TO DISCONNECT SURGEON SIDE CONSOLE (SSC)1) AND RELEASE THE TISSUE GRASPED BY THE FORCE BIPOLAR FORCEPS INSTRUMENT. AFTER FOLLOWING THESE STEPS, THE SYSTEM BOOTED UP FINE, AND THE CUSTOMER WAS ABLE TO PROCEED WITH THE SURGERY USING THE OTHER SSC. A PROCEDURE DELAY WAS REPORTED. NO KNOWN IMPACT OR PATIENT CONSEQUENCE WAS REPORTED.
REFER TO H11 FOR FOLLOW-UP INFORMATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 149896 | DAVINCI XI | SURGEON SIDE CONSOLE, SMART PEDALS | NAY | INTUITIVE SURGICAL, INC | 380677-28 | N/A | 00886874110744 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown | DA VINCI INSTRUMENTS AND ACCESSORIES. |