FDA Adverse Event Injury Summary report: N

STYLE 68 SALINE FILLED BREAST IMPLANT

MDR report key: 24512105 · Received March 4, 2026

Report

Report Number
9617229-2026-03969
Event Type
Injury
Date Received
March 4, 2026
Date of Event
December 16, 2025
Report Date
March 4, 2026
Manufacturer
ALLERGAN (COSTA RICA)
Product Code
FWM
UDI-DI
10888628000247
PMA / PMN Number
P990074
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

ADDITIONAL, CHANGED, AND/OR CORRECTED DATA: D4(UDI) D9 H3 H6. LABORATORY ANALYSIS SUMMARY: THE DEVICE RELATED TO THE REPORTED EVENTS DEFLATION WAS RECEIVED ON APRIL 29, 2026 WITH LOT NUMBER 2163349. PER THE INVESTIGATION PROCEDURE, THE DEVICE IS ANALYZED THROUGH VISUAL INSPECTION MICROSCOPIC INSPECTION IF OPENINGS ARE OBSERVED AND A WEIGHT VERIFICATION. PER THE ANALYSIS PERFORMED, THE ASSESSMENTS OF THE COMPLAINTS AND ANY POTENTIAL MANUFACTURING ISSUE ARE DISPLAYED ALONG WITH ANY FURTHER ACTIONS REQUIRED: - DEFLATION: OBSERVED AN OPENING THROUGH MICROSCOPIC INSPECTION ASSESSED AS FOLD FLAW OPENING. NO FURTHER ACTIONS ARE REQUIRED AS IT IS NOT RELATED TO THE MANUFACTURING PROCESS. NONE OF THE OTHER OBSERVATIONS PERFORMED DURING THE DEVICE ANALYSIS CREASES AND WEAR ABRASION ARE FOUND TO BE POTENTIALLY RELATED TO THE MANUFACTURING PROCESS, AND, THEREFORE, NO FURTHER ACTIONS ARE REQUIRED FOR THESE OBSERVATIONS.

Additional Manufacturer Narrative · 0

A REVIEW OF THE DEVICE HISTORY RECORD HAS BEEN COMPLETED. NO DEVIATIONS OR NON-CONFORMANCE'S NOTED. FURTHER INFORMATION FROM THE REPORTER REGARDING EVENT, PRODUCT, OR PATIENT DETAILS HAS BEEN REQUESTED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME. REASON FOR REOPERATION: DEFLATION.

Description of Event or Problem · 0

HEALTHCARE PROFESSIONAL REPORTED "DEFLATION". THIS RECORD IS FOR THE RIGHT SIDE. THE DEVICE WAS EXPLANTED.

Description of Event or Problem · 0

HEALTHCARE PROFESSIONAL REPORTED "DEFLATION". THIS RECORD IS FOR THE RIGHT SIDE. THE DEVICE WAS EXPLANTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
567889 STYLE 68 SALINE FILLED BREAST IMPLANT PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE FWM ALLERGAN (COSTA RICA) 2163349 10888628000247

Patients

Seq Age Sex Outcome Treatment
1 42 YR Female Required Intervention