FDA Adverse Event Malfunction Summary report: N

MINIMED 780G US SYSTEM BLE CONNECT 3.0 MG/DL

MDR report key: 24493422 · Received March 3, 2026

Report

Report Number
2032227-2026-137590
Event Type
Malfunction
Date Received
March 3, 2026
Date of Event
February 18, 2026
Report Date
March 3, 2026
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OZP
UDI-DI
000000763000924089
PMA / PMN Number
P160017
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NE, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

THIS MDR RELATED TO THE PUERTO RICO MANUFACTURING SITE HAS BEEN ASSIGNED A MEDWATCH NUMBER FROM THE MEDTRONIC MINIMED NORTHRIDGE SITE, PER VARIANCE 5. CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. WE THEREFORE CONSIDER THIS REPORT COMPLETE TO THE BEST OF OUR KNOWLEDGE. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Description of Event or Problem · 0

ON 02/19/2026 16:43:00 (B)(4): BATTERY FAILED ALARM CUSTOMER REPORTED RECEIVING A BATTERY FAILED ALARM. 02/19/2026 16:43:00 (B)(4) GENERAL DOCUMENTATION THE CUSTOMER CALLED FOR AN ISSUE NOT COVERED BY EXISTING TROUBLESHOOTING GUIDES. REFER TO THE EVENT DESCRIPTION FOR MORE DETAILS. 02/19/2026 16:43:00 (B)*=(4): RESERVOIR <(>&<)> TUBING/SET/PRIMING PROCESS TROUBLESHOOTING CUSTOMER REPORTED INSULIN SQUIRTING OUT/DRIPPING OUT DURING FILL TUBING PROCESS. 02/19/2026 16:43:00 (B)(4): BUMPED/DROPPED/COSMETIC DAMAGE CUSTOMER REPORTED PUMP WAS DROPPED/BUMPED AND/OR PUMP HAS POSSIBLE PHYSICAL OR COSMETIC DAMAGE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
556396 MINIMED 780G US SYSTEM BLE CONNECT 3.0 MG/DL AUTOMATED INSULIN DOSING DEVICE SYSTEM, SINGLE HORMONAL CONTROL OZP MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-1884 NG4023662H 000000763000924089

Patients

Seq Age Sex Outcome Treatment
1 58 YR Male