SPRINT QUATTRO SECURE
Report
- Report Number
- 2649622-2012-00426
- Event Type
- Injury
- Date Received
- February 13, 2012
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- LWS
- PMA / PMN Number
- P920015/S24
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- PHYSICIAN
Narratives
THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. EVALUATION SUMMARY: (B)(4): THE ACTUAL LEAD WAS NOT RECEIVED FOR EVALUATION. WE DID RECEIVE PERFORMANCE DATA COLLECTED FROM THE DEVICE AND HAVE ANALYZED THE DATA. ONE -PATIENT ALERT FOR OUT OF TOLERANCE SUBTHRESHOLD LEAD IMPEDANCE ON (B)(4) 2011 02:15:03. WEEKLY PACE LEAD IMPEDANCE TREND DATA SHOWS AN ABRUPT INCREASE FOR MAX RV PACE = 592 TO INFINITE OHMS (UN-FILTERED DATA) PEAK BETWEEN (B)(4) 2011 AND (B)(4) 2011. ELEVEN - VENTRICULAR NST (NON-SUSTAINED TACHYCARDIA) <=210 MS BETWEEN (B)(4) 2011 14:00:03 AND (B)(4) 2011 16:56:40. ONE - VF=180 MS AVERAGE V-CYCLE (B)(4) 2011 10:14:06. VENTRICULAR SHORT INTERVAL COUNT V-SIC=421.4 COUNTS AVG/DAY, IN 20.74 DAYS, BETWEEN (B)(4) 2011 16:39:47 AND (B)(4) 2011 10:19:40.
IT WAS REPORTED THAT DURING A ROUTINE FOLLOW-UP, HIGH PACING IMPEDANCE GREATER THAN 2500 OHMS WAS NOTED, ALONG WITH OVER 8000 SHORT INTERVALS WHICH COULD INDICATE OVERSENSING. THE LEAD WAS REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SPRINT QUATTRO SECURE | IMPLANTABLE TACHY LEAD | LWS | MEDTRONIC PUERTO RICO, INC. | 6947 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Hospitalization| R |