FDA Adverse Event Malfunction Summary report: N

ALINITY I STAT HIGH SENSITIVE TROPONIN-I REAGENT KIT

MDR report key: 24484472 · Received March 2, 2026

Report

Report Number
3005094123-2026-00102
Event Type
Malfunction
Date Received
March 2, 2026
Date of Event
February 12, 2026
Report Date
May 21, 2026
Manufacturer
ABBOTT IRELAND DIAGNOSTICS DIVISION
Product Code
MMI
UDI-DI
00380740160647
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IS
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

SECTION A PATIENT INFORMATION: REFER TO SECTION B5 FOR COMPLETE PATIENT INFORMATION. THIS REPORT IS BEING FILED ON AN INTERNATIONAL PRODUCT, LIST NUMBER 08P13-34 THAT HAS A SIMILAR PRODUCT DISTRIBUTED IN THE US, ALINITY I STAT HIGH SENSITIVITY TROPONIN-I, LIST NUMBER 04Z21, WITH 510K/PMA/BLA NUMBER K202525. AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE.

Description of Event or Problem · 0

THE CUSTOMER OBSERVED FALSELY ELEVATED ALINITY I STAT HIGH SENSITIVE TROPONIN-I RESULTS GENERATED BY THE ALINITY I PROCESSING MODULE FOR MULTIPLE PATIENTS. THE FOLLOWING DATA WERE PROVIDED: LAB REFERENCE RANGE: FEMALE IS <16 NG/L, MALE IS <34 NG/L LOT#80170UD00 ON INSTRUMENT SERIAL NUMBER (B)(6): (B)(6) 2026, SID (B)(6), (MALE) INITIAL RESULT 29.26 NG/L, REPEAT RESULT = 37.59 NG/L (B)(6) 2026, SID (B)(6), (FEMALE) INITIAL RESULT 14.86 NG/L, REPEAT RESULT = 16.88 NG/L (B)(6) 2026, SID (B)(6), (MALE) INITIAL RESULT 47.58 NG/L, REPEAT RESULT = 29.6 NG/L (B)(6) 2026, SID (B)(6), (FEMALE) INITIAL RESULT 11.42 NG/L, REPEAT RESULT = 16.74 NG/L (B)(6) 2026, SID (B)(6), (FEMALE) INITIAL RESULT 16.9 NG/L, REPEAT RESULT = 15.21 NG/L (B)(6) 2026, SID (B)(6), (MALE) INITIAL RESULT 35.21 NG/L, REPEAT RESULT = 29.02 NG/L NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
346290 ALINITY I STAT HIGH SENSITIVE TROPONIN-I REAGENT KIT IMMUNOASSAY METHOD, TROPONIN SUBUNIT MMI ABBOTT IRELAND DIAGNOSTICS DIVISION 80170UD00 00380740160647

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown ALNTY I PROCESSING MODU, 03R65-01,(B)(6)