FDA Adverse Event Malfunction Summary report: N

ALINITY I STAT HIGH SENSITIVE TROPONIN-I REAGENT KIT

MDR report key: 24484452 · Received March 2, 2026

Report

Report Number
3005094123-2026-00101
Event Type
Malfunction
Date Received
March 2, 2026
Date of Event
February 2, 2026
Report Date
May 21, 2026
Manufacturer
ABBOTT IRELAND DIAGNOSTICS DIVISION
Product Code
MMI
UDI-DI
00380740160647
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IS
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

SECTION A PATIENT INFORMATION: REFER TO SECTION B5 FOR COMPLETE PATIENT INFORMATION. THIS REPORT IS BEING FILED ON AN INTERNATIONAL PRODUCT, LIST NUMBER 08P13-34 THAT HAS A SIMILAR PRODUCT DISTRIBUTED IN THE US, ALINITY I STAT HIGH SENSITIVITY TROPONIN-I, LIST NUMBER 04Z21, WITH 510K/PMA/BLA NUMBER K202525. AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE.

Description of Event or Problem · 0

THE CUSTOMER OBSERVED FALSELY ELEVATED ALINITY I STAT HIGH SENSITIVE TROPONIN-I RESULTS GENERATED BY THE ALINITY I PROCESSING MODULE FOR MULTIPLE PATIENTS. THE FOLLOWING DATA WERE PROVIDED: LAB REFERENCE RANGE: FEMALE IS <16 NG/L, MALE IS <34 NG/L. LOT#79329UD00: (B)(6) 2026, SID (B)(6), (72-YEAR-OLD, FEMALE): INITIAL RESULT = 242 NG/L (B)(6), REPEAT RESULT = <10 NG/L (B)(6). (B)(6) 2026, SID (B)(6), (52-YEAR-OLD, MALE): INITIAL RESULT = 59.36 NG/L (B)(6), REPEAT RESULT = <10 NG/L (B)(6). (B)(6) 2026, SID (B)(6), (70-YEAR-OLD, MALE): INITIAL RESULT = 65.46 NG/L (B)(6), REPEAT RESULT = 23.55 NG/L (B)(6). NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
68721 ALINITY I STAT HIGH SENSITIVE TROPONIN-I REAGENT KIT IMMUNOASSAY METHOD, TROPONIN SUBUNIT MMI ABBOTT IRELAND DIAGNOSTICS DIVISION 79329UD00 00380740160647

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown ALNTY I PROCESSING MODU, 03R65-01, (B)(6),| ALNTY I PROCESSING MODU, 03R65-01, (B)(6).