FDA Adverse Event Malfunction Summary report: N

DAVINCI XI

MDR report key: 24478691 · Received March 2, 2026

Report

Report Number
2955842-2026-14671
Event Type
Malfunction
Date Received
March 2, 2026
Date of Event
January 28, 2026
Report Date
May 19, 2026
Manufacturer
INTUITIVE SURGICAL, INC
Product Code
NAY
UDI-DI
00886874110744
PMA / PMN Number
K131861
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

AN INTUITIVE SURGICAL, INC. (ISI) FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO THE CUSTOMER SITE TO FURTHER INVESTIGATE THE REPORTED EVENT. FSE REPLACED THE HELICAL COUPLER AND GAS SPRINGS TO CORRECT REPORTED PROBLEM WITH THE FOOTREST LIFT. THE SYSTEM WAS TESTED AND VERIFIED AS READY FOR USE. THE AFFECTED PART INVOLVED WITH THIS COMPLAINT IS A FIELD SCRAP ITEM AND WILL NOT BE RETURNED TO ISI FOR FURTHER INVESTIGATION. THE COMPLAINT WAS CONFIRMED BASED ON THE FIELD EVALUATION, WHICH INDICATES THAT THE DEVICE MAY HAVE CONTRIBUTED TO THE CUSTOMER REPORTED ISSUE. THE PROBABLE ROOT CAUSE COULD NOT BE DETERMINED.

Additional Manufacturer Narrative · 0

THE ROOT CAUSE IS ATTRIBUTED TO A FAULTY HELICAL COUPLER AND THE GAS SPRING CAUSING MOVEMENT ISSUES IN THE FOOT TRAY.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED SURGICAL PROCEDURE, THE CUSTOMER CALLED AN ISI TECHNICAL SUPPORT ENGINEER (TSE) AND REPORTED THE SITE STATED THEY WERE NOT ABLE TO ADJUST THE FOOT PEDALS ON THE SURGEON SIDE CONSOLE (SSC). SITE COULD HEAR THE MOTOR WITH NOTHING MOVING. SITE HAD CHANGED OUT SSC PRIOR TO CALLING IN TO COMPLETE THE CASE. THE PROCEDURE WAS CONVERTED TO ANOTHER DV AND COMPLETED WITH NO REPORTED INJURY. ISI FOLLOWED UP WITH THE INITIAL REPORTER AND OBTAINED THE FOLLOWING ADDITIONAL INFORMATION: THE PORTS WERE NOT PLACED ON THE PATIENT. IT WAS ABORTED.

Description of Event or Problem · 0

REFER TO H11 FOR FOLLOW-UP INFORMATION.