FDA Adverse Event Injury Summary report: N

SPECTRA WAVEWRITER?

MDR report key: 24477783 · Received March 2, 2026

Report

Report Number
3006630150-2026-01178
Event Type
Injury
Date Received
March 2, 2026
Date of Event
July 15, 2025
Report Date
April 22, 2026
Manufacturer
BOSTON SCIENTIFIC NEUROMODULATION CORPORATION
Product Code
LGW
UDI-DI
08714729951254
PMA / PMN Number
P030017
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE (IPG AND LEADS) WERE NOT RETURNED FOR ANALYSIS; THEREFORE, A TECHNICAL ANALYSIS COULD NOT BE PERFORMED. IT WAS CONFIRMED THESE DEVICES MET MANUFACTURING SPECIFICATION PRIOR TO DISTRIBUTION AND THERE WERE NO MANUFACTURING DEVIATIONS WHICH COULD HAVE CONTRIBUTED TO THE REPORTED EVENT. REVIEW OF THE INSTRUCTIONS FOR USE (IFU) CONFIRMED THERE WAS RELEVANT CONTENT AND SUFFICIENT GUIDANCE WITH RESPECT TO THE CIRCUMSTANCES DESCRIBED WITHIN THIS COMPLAINT. NO UPDATES ARE REQUIRED TO THE IFU AS A RESULT OF THIS EVENT. BASED ON A THOROUGH REVIEW OF THE REPORTED COMPLAINT, BOSTON SCIENTIFIC HAS ASSIGNED AN INVESTIGATION CONCLUSION CODE OF KNOWN INHERENT RISK OF DEVICE.

Additional Manufacturer Narrative · 0

ADDITIONAL SUSPECT MEDICAL DEVICE COMPONENTS INVOLVED IN THE EVENT: PRODUCT FAMILY: SCS-LINEAR LEADS. UPN: M365SC2317500. MODEL: SC-2317-50. SERIAL: (B)(6). BATCH: 5031269. UDI: (B)(4). PRODUCT FAMILY: SCS-LINEAR LEADS. UPN: M365SC2317500. MODEL: SC-2317-50. SERIAL: (B)(6). BATCH: 21562676. UDI: (B)(4). PRODUCT FAMILY: SCS-LEAD FIXATION. UPN: M365SC43180. MODEL: SC-4318. BATCH: 21738704. UDI: (B)(4).

Description of Event or Problem · 0

IT WAS REPORTED THAT THE PATIENT WAS EXPERIENCING INADEQUATE STIMULATION DUE TO HIGH IMPEDANCES AND LEAD MIGRATION CONFIRMED THROUGH X RAY IMAGING. THE PATIENT UNDERWENT A SPINAL CORD STIMULATOR (SCS) SYSTEM REPLACEMENT PROCEDURE, DURING WHICH A PADDLE LEAD WAS IMPLANTED. THE PATIENT WAS DOING WELL POSTOPERATIVELY, AND ALL EXPLANTED DEVICES WERE NOT RETURNED DUE TO FACILITY POLICY.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE PATIENT WAS EXPERIENCING INADEQUATE STIMULATION DUE TO HIGH IMPEDANCES AND LEAD MIGRATION CONFIRMED THROUGH X RAY IMAGING. THE PATIENT UNDERWENT A SPINAL CORD STIMULATOR (SCS) SYSTEM REPLACEMENT PROCEDURE, DURING WHICH A PADDLE LEAD WAS IMPLANTED. THE PATIENT WAS DOING WELL POSTOPERATIVELY, AND ALL EXPLANTED DEVICES WERE NOT RETURNED DUE TO FACILITY POLICY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
171436 SPECTRA WAVEWRITER? STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF LGW BOSTON SCIENTIFIC NEUROMODULATION CORPORATION SC-1160 333405 08714729951254

Patients

Seq Age Sex Outcome Treatment
1 61 YR Female Required Intervention