Report

Report Number
3005094123-2026-00099
Event Type
Malfunction
Date Received
February 26, 2026
Date of Event
February 16, 2026
Report Date
April 29, 2026
Manufacturer
ABBOTT IRELAND DIAGNOSTICS DIVISION
Product Code
MMI
UDI-DI
00380740118075
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IT
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE. SECTION A PATIENT INFORMATION: REFER TO SECTION B5 FOR COMPLETE PATIENT INFORMATION. ALL AVAILABLE PATIENT INFORMATION WAS PROVIDED, NO ADDITIONAL PATIENT INFORMATION WAS AVAILABLE. THIS REPORT IS BEING FILED ON AN INTERNATIONAL PRODUCT, LIST NUMBER: 3P25, THAT HAS A SIMILAR PRODUCT DISTRIBUTED IN THE US, LIST NUMBER: 2R98, AND A 510K/PMA/BLA NUMBER: K191595.

Description of Event or Problem · 0

THE CUSTOMER OBSERVED FALSELY ELEVATED ARCHITECT STAT HIGHLY SENSITIVE TROPONIN-I RESULTS GENERATED FOR A MALE PATIENT SAMPLE. THE FOLLOWING DATA WAS PROVIDED (CUSTOMER¿S REFERENCE RANGE FOR MALES 34.2 PG/ML): ON (B)(6) 2026, SAMPLE ID: (B)(6) INITIAL RESULT = 44.9 PG/ML, REPEAT RESULTS = 4 PG/ML, 4.3 PG/ML. SAME PATIENT NEW SAMPLE OBTAINED. SAMPLE ID: (B)(6) RESULT = 4.7 PG/ML, REPEAT RESULT = 4 PG/ML NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
85859 ARCHITECT STAT HIGH SENSITIVE TROPONIN-I IMMUNOASSAY METHOD, TROPONIN SUBUNIT MMI ABBOTT IRELAND DIAGNOSTICS DIVISION 80401UD00 00380740118075

Patients

Seq Age Sex Outcome Treatment
1 NA Male ARC I1000SR MOD, 01L86-01, I1SR01771| ARC I1000SR MOD, 01L86-01, I1SR01771