FDA Adverse Event Malfunction Summary report: N

MINIMED 780G US SYSTEM BLE CONNECT 3.0 MG/DL

MDR report key: 24454618 · Received February 26, 2026

Report

Report Number
2032227-2026-135429
Event Type
Malfunction
Date Received
February 26, 2026
Date of Event
February 9, 2026
Report Date
April 20, 2026
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OZP
PMA / PMN Number
P160017
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

UNIT PASSED DISPLACEMENT TEST, REWIND TEST, PRIME/SEATING TEST, BASIC OCCLUSION TEST, FORCE SENSOR TEST, AND SELF-TEST. UNIT SUCCESSFULLY DOWNLOADED TO THUMP. NO PUMP ERROR 15 AND 23 NOTED DURING TESTING. HOWEVER, CONFIRMED PUMP ERROR 15 (LINE NUMBER 442 FILE NUMBER 38) IN THE PUMP HISTORY DOWNLOAD ON 02/09/2026 19:18:09 DUE TO SW ISSUE AS PER GLOBAL LOGIC ANALYSIS. VERIFIED THE PUMP ALARMED PUMP ERROR 23 AFTER BATTERY REMOVAL ON 02/09/2026 19:18:11 IN PUMP DOWNLOADED HISTORY. THESE ALERTS ARE EXPECTED AND OCCUR DURING NORMAL PUMP OPERATION. THE ELECTRONIC ASSEMBLIES AND MOTOR ASSEMBLIES WERE INSPECTED, AND NO ANOMALIES WERE NOTED. THE P-CAP LOCKS PROPERLY INTO THE RESERVOIR COMPARTMENT. THE FOLLOWING WERE NOTED DURING VISUAL INSPECTION: SCRATCHED CASE, PILLOWING KEYPAD OVERLAY, CRACKED CASE CORNER OF THE BELT CLIP RAILS NEAR THE BATTERY COMPARTMENT, AND CRACKED KEYPAD OVERLAY. PUMP PASSED ALL REQUIRED TESTING, AND NO PUMP ERROR 23 NOTED DURING ANALYSIS. CONFIRMED PUMP ERROR 15 IN THE PUMP HISTORY DOWNLOAD DUE TO SW ISSUE AS PER GLOBAL LOGIC ANALYSIS. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Additional Manufacturer Narrative · 0

THIS MDR RELATED TO THE PUERTO RICO MANUFACTURING SITE HAS BEEN ASSIGNED A MEDWATCH NUMBER FROM THE MEDTRONIC MINIMED NORTHRIDGE SITE, PER VARIANCE 5. CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Description of Event or Problem · 0

IT WAS REPORTED TO MEDTRONIC MINIMED THAT THE CUSTOMER EXPERIENCED HYPOGLYCEMIA. THE CUSTOMER ALSO REPORTED PUMP ERROR 15(THE CRITICAL BLOCK AND INVERSE CRITICAL BLOCK DID NOT ADD TO ZERO), PUMP ERROR 23(POST-RESET RAM CRC ALARM). THE EVENT INVOLVED PRODUCT(S) MMT-1884. TROUBLESHOOTING WAS PERFORMED. THE ALARM WAS CLEARED, THE PUMP WAS REWOUND, FILL CANNULA AND SELF-TESTS WERE COMPLETED SUCCESSFULLY. IT IS UNKNOWN WHETHER THE CUSTOMER WAS USING THE AUTO MODE/SMART GUARD FEATURE AT THE TIME OF THE EVENT. IT IS UNKNOWN WHETHER THE CUSTOMER WAS USING THE INSULIN PUMP SYSTEM WITHIN 48 HOURS OF REPORTED EVENT. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED. THE CUSTOMER WILL DISCONTINUE USE OF THE INSULIN PUMP. MMT-1884 WAS REQUESTED AND THE CUSTOMER RESPONSE WAS THAT THE DEVICE WILL BE RETURNED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
398397 MINIMED 780G US SYSTEM BLE CONNECT 3.0 MG/DL AUTOMATED INSULIN DOSING DEVICE SYSTEM, SINGLE HORMONAL CONTROL OZP MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-1884 NG3745450H

Patients

Seq Age Sex Outcome Treatment
1 68 YR Male