Report
- Report Number
- 2955842-2026-04872
- Event Type
- Malfunction
- Date Received
- February 25, 2026
- Date of Event
- February 2, 2026
- Report Date
- April 10, 2026
- Manufacturer
- INTUITIVE SURGICAL, INC
- Product Code
- NAY
- UDI-DI
- 00886874116555
- PMA / PMN Number
- K191736
- Removal / Correction Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN
- Reporter Occupation
- BIOMEDICAL ENGINEER
- Health Professional
- Yes
Narratives
INTUITIVE SURGICAL INC. (ISI) RECEIVED THE 30-DEGREE ENDOSCOPE PLUS FOR FAILURE ANALYSIS INVESTIGATION. THE ENDOSCOPE WAS ANALYZED AND PLACED THROUGH FRICTION TEST USING A SCREENING AID TO MANUALLY ROTATE THE ADAPTER TO DETECT FOR FRICTION. THE CAMERA INSTRUMENT ADAPTER DID NOT ROTATE PROPERLY AND/OR NOISES WERE HEARD DURING TESTING AND CONFIRMED AS BINDING. ADDITIONALLY, THE ENDOSCOPE WAS TESTED AND PLACED ON A CAMERA ATTENUATION TESTER TO MEASURE TRANSMITTANCE, BUT FAILED FUNCTIONAL TESTING. THE ENDOSCOPE FAILED TRANSMITTANCE DUE TO THE MEASURED OUTPUT POWER NOT MEETING SPECIFICATION LIMITS. THE COMPLAINT WAS CONFIRMED BY FAILURE ANALYSIS. THE PROBABLE ROOT CAUSE OF BINDING IN CAMERA INSTRUMENT ADAPTER IS ATTRIBUTED TO COMPONENT SUSCEPTIBILITY TO INCREASED FRICTION.
INTUITIVE SURGICAL, INC. (ISI) HAS NOT RECEIVED THE DA VINCI PRODUCT WITH AN ALLEGED ISSUE TO PERFORM FAILURE ANALYSIS. A FOLLOW-UP MDR WILL BE SUBMITTED IF THE PRODUCT IS RETURNED AND EVALUATED AND/OR IF ADDITIONAL INFORMATION IS RECEIVED. BLANK MDR FIELDS: THE MISSING PATIENT INFORMATION IN SECTIONS A AND B WAS EITHER UNKNOWN, UNAVAILABLE, NOT PROVIDED, OR NOT APPLICABLE. THE EXPIRATION DATE FOR SECTION D4 IS NOT APPLICABLE. FIELD D6 IS BLANK BECAUSE THE PRODUCT IS NOT IMPLANTABLE. INFORMATION FOR THE BLANK FIELDS IN SECTION E1 IS NOT AVAILABLE. FIELDS G5 AND G7 ARE NOT APPLICABLE. FIELD H10 IS BLANK AS THERE ARE NO RELATED REPORT NUMBERS.
REFER TO H11 FOR FOLLOW-UP INFORMATION.
IT WAS REPORTED THAT DURING DA VINCI-ASSISTED BILATERAL INTERNAL MAMMARY ARTERY HARVEST SURGICAL PROCEDURE, SITE CONTACTED INTUITIVE TECHNICAL SUPPORT ENGINEER (TSE) TO REPORT ROTATION ISSUE WITH THE 30-DEGREE ENDOSCOPE PLUS. THE SITE COULD NOT USE IT. THE PROCEDURE WAS COMPLETED WITH NO REPORTED INJURY. INTUITIVE SURGICAL, INC. (ISI) FOLLOWED UP WITH THE INITIAL REPORTER AND OBTAINED THE FOLLOWING ADDITIONAL INFORMATION: THE LOGS SHOW THAT THE SCOPE WAS USED ON (B)(6) 2026.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 372361 | NONE | ENDOSCOPE PLUS | NAY | INTUITIVE SURGICAL, INC | 470057-08 | N/A | 00886874116555 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown | DA VINCI INSTRUMENTS AND ACCESSORIES |