ARCHITECT TOTAL PSA
Report
- Report Number
- 3008344661-2026-00044
- Event Type
- Malfunction
- Date Received
- February 25, 2026
- Date of Event
- February 11, 2026
- Report Date
- March 19, 2026
- Manufacturer
- ABBOTT IRELAND DIAGNOSTICS DIVISION
- Product Code
- MTF
- UDI-DI
- 00380740112417
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE. ALL AVAILABLE PATIENT INFORMATION HAS BEEN INCLUDED. NO ADDITIONAL PATIENT DETAILS ARE AVAILABLE. SECTION A PATIENT INFORMATION: REFER TO SECTION B5 FOR COMPLETE PATIENT INFORMATION. THIS REPORT IS BEING FILED ON AN INTERNATIONAL PRODUCT, LIST NUMBER 07K70 THAT HAS A SIMILAR PRODUCT DISTRIBUTED IN THE US, LIST NUMBER 6C06, WITH 510K NUMBER P910007.
ADDITIONAL INFORMATION/CORRECTION: AFTER FURTHER EVALUATION, THE SUSPECT MEDICAL DEVICE WAS CHANGED FROM ARCHITECT TOTAL PSA, LIST 07K70-35, MANUFACTURING SITE SLIGO, TO THE ARCHITECT I2000SR PROCESSING MODULE, 03M74-01 MANUFACTURING SITE DALLAS. MDR NUMBER 3016438761-2026-00149 HAS BEEN SUBMITTED AND ALL FURTHER INFORMATION WILL BE DOCUMENTED UNDER THAT MDR NUMBER.
THE CUSTOMER OBSERVED FALSELY ELEVATED ARCHITECT TOTAL PSA RESULTS FOR ONE PATIENT. THE FOLLOWING DATA WAS PROVIDED: SID UNKNOWN, INITIAL PSA RESULT WAS 6.18 NG/ML, REPEATED 0.11 NG/ML. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
THE CUSTOMER OBSERVED FALSELY ELEVATED ARCHITECT TOTAL PSA RESULTS FOR ONE PATIENT. THE FOLLOWING DATA WAS PROVIDED: SID UNKNOWN, INITIAL PSA RESULT WAS 6.18 NG/ML, REPEATED 0.11 NG/ML. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 503037 | ARCHITECT TOTAL PSA | TOTAL, PROSTATE SPECIFIC AG (NONCOMPLEXED/COMPLEXED) FOR DET OF PROSTATE CANCER | MTF | ABBOTT IRELAND DIAGNOSTICS DIVISION | 78463FZ00 | 00380740112417 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |