ALINITY I STAT HIGH SENSITIVE TROPONIN-I REAGENT KIT
Report
- Report Number
- 3005094123-2026-00093
- Event Type
- Malfunction
- Date Received
- February 25, 2026
- Date of Event
- February 14, 2026
- Report Date
- April 22, 2026
- Manufacturer
- ABBOTT IRELAND DIAGNOSTICS DIVISION
- Product Code
- MMI
- UDI-DI
- 00380740160647
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE. SECTION A PATIENT INFORMATION: REFER TO SECTION B5 FOR COMPLETE PATIENT INFORMATION. ALL AVAILABLE PATIENT INFORMATION WAS INCLUDED. ADDITIONAL PATIENT DETAILS ARE NOT AVAILABLE. THIS REPORT IS BEING FILED ON AN INTERNATIONAL PRODUCT, LIST NUMBER 08P13-34, THAT HAS A SIMILAR PRODUCT DISTRIBUTED IN THE US, LIST NUMBER 04Z21, ALINITY I STAT HIGH SENSITIVITY TROPONIN-I, WITH 510K/PMA/BLA NUMBER K202525.
THE CUSTOMER OBSERVED FALSE ELEVATED ALINITY I STAT HIGH SENSITIVE TROPONIN-I GENERATED FROM ALINITY I PROCESSING MODULE (SN: (B)(6) AND PROVIDED THE FOLLOWING DATA: CUSTOMER REFERENCE RANGE: 0 - 4.9 NG/L PATIENT 1: 25 YEAR OLD FEMALE SAMPLE ID (B)(6). ON (B)(6) 2026, INITIAL RESULT WAS 7.0 NG/L. ON (B)(6) 2026, THE SAMPLE WAS REPEATED ON ANOTHER ANALYZER (SN: (B)(6) AND THE RESULT WAS 1.13 NG/L. PATIENT 2: 56 YEAR OLD MALE SAMPLE ID (B)(6). ON (B)(6) 2026, INITIAL RESULT WAS 13.4 NG/L. WHEN REPEATED ON ANOTHER ANALYZER (SN: (B)(6) AND THE RESULT WAS 1.7 NG/L. PER THE CUSTOMER DISCREPANT RESULTS WERE NOT REPORTED OUT TO THE MEDICAL PROVIDER. THERE WAS NO REPORTED IMPACT TO PATIENT MANAGEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 510403 | ALINITY I STAT HIGH SENSITIVE TROPONIN-I REAGENT KIT | IMMUNOASSAY METHOD, TROPONIN SUBUNIT | MMI | ABBOTT IRELAND DIAGNOSTICS DIVISION | 81566UD00 | 00380740160647 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown | ALNTY I PROCESSING MODU, 03R65-01, (B)(6). |