I-STAT TROPONIN (HS-TNI ) CARTRIDGE
Report
- Report Number
- 2245578-2026-00045
- Event Type
- Malfunction
- Date Received
- February 24, 2026
- Date of Event
- February 2, 2026
- Report Date
- March 18, 2026
- Manufacturer
- ABBOTT POINT OF CARE INC.
- Product Code
- MMI
- PMA / PMN Number
- K240984
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
APOC INCIDENT # (B)(4). APOC LABELING WILL BE EVALUATED DURING THE INVESTIGATION AS PERTAINING TO THE EVENT.
APOC INCIDENT: (B)(4). THE INVESTIGATION WAS COMPLETED ON 12-MAR-2026. A REVIEW OF THE DEVICE HISTORY RECORD (DHR) CONFIRMED THE CARTRIDGE LOT MET FINISHED GOODS (FG) RELEASE CRITERIA. A DEFICIENCY HAS BEEN IDENTIFIED FOR HS-TNI CARTRIDGE LOT A25210B. WHILE RETAINED CARTRIDGE TESTING MET THE ACCEPTANCE CRITERIA FOR THE DETECTED ERROR (DT) AND UNDETECTED ERROR (UDT) RATES OUTLINED IN APPENDIX 1 OF Q04.01.003 REV. AP (PRODUCT COMPLAINT LEVEL 2 AND LEVEL 3 INVESTIGATION PROCEDURE), THE FG RELEASE SPECIFICATION FOR STANDARD DEVIATION (SD) PER TP-0651 REV. BI (CARTRIDGE FINISHED GOODS TEST SPECIFICATION) WAS NOT MET. ROOT CAUSE WILL BE INVESTIGATED IN QUALITY RECORD (QR) 1100932.
ON 19-FEB-2026, ABBOTT POINT OF CARE WAS CONTACTED BY A CUSTOMER REGARDING I-STAT TROPONIN CARTRIDGES THAT YIELDED DISCREPANT FALSE POSITIVE RESULTS ON A 46 YEAR OLD MALE PATIENT WITH CHEST PAIN & SUBACUTE COUGH. THERE WAS NO ADDITIONAL PATIENT INFORMATION. RETURN PRODUCT IS AVAILABLE FOR INVESTIGATION. METHOD DATE COLLECTED RESULT SAMPLE POSITIVE/NEGATIVE I-STAT (B)(6) 2026 14:47 976 A POSITIVE ROCHE (B)(6) 2026 14:47 11 A NEGATIVE I-STAT (B)(6) 2026 16:44 635 B POSITIVE ROCHE (B)(6) 2026 16:44 7 B NEGATIVE I-STAT (B)(6) 2026 17:37 590 C POSITIVE ROCHE (B)(6) 2026 17:37 9 C NEGATIVE. THERE ARE NO INJURIES ASSOCIATED WITH THIS EVENT. AT THIS TIME THERE IS NO REASON TO SUSPECT A MALFUNCTION EXISTS. THE INVESTIGATION IS UNDERWAY. 99TH PERCENTILE 90% CI SEX N (NG/L, PG/ML) (NG/L, PG/ML) FEMALE 490 13 (10, 17) MALE 404 28 (19, 58) OVERALL 895 21 (14, 30).
NA.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 3179 | I-STAT TROPONIN (HS-TNI ) CARTRIDGE | HS-TNI CARTRIDGE | MMI | ABBOTT POINT OF CARE INC. | NA | A25210B |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 46 YR | Male |