FDA Adverse Event Malfunction Summary report: N

SIMVIEW 3000 SIMULATOR

MDR report key: 2443390 · Received January 19, 2012

Report

Report Number
2910081-2012-00006
Event Type
Malfunction
Date Received
January 19, 2012
Date of Event
December 20, 2011
Report Date
December 20, 2011
Manufacturer
SIEMENS MEDICAL SOLUTIONS USA, INC., RADIATION ONCOLOGY
Product Code
KPQ
PMA / PMN Number
K946053
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PL
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

SIEMENS BECAME AWARE OF THE REPORTED EVENT ON (B)(4) 2011 AND ARE MAILING THE MDR ON (B)(4) 2012. SIEMENS CUSTOMER SERVICE ENGINEER FOUND DURING EVALUATION THAT THE MAIN CHAIN FOR THE MOTOR AND THE LIFTING SCREW OR BARRING WERE BOTH BROKEN. THE BROKEN CHAIN WAS REPLACED, AND A NEW TABLE BASE AND TABLE CABLE WERE INSTALLED. THE SIMULATOR IS BACK IN USE BY THE CUSTOMER. (B)(6).

Description of Event or Problem · 1

SIEMENS RECEIVED A REPORT ON (B)(6) 2011, NOTING THAT WHEN THE OPERATOR MOVED THE TABLE DOWN AT THE END OF A PATIENT'S SIMULATION SESSION ON THE SIMVIEW 3000 SIMULATOR, THE TABLE RAPIDLY FELL DOWN ABOUT 50 CM WITH THE PATIENT ON TOP OF IT. REPORTEDLY, THE PATIENT WAS NOT INJURED. THE PATIENT CAME DOWN FROM THE TABLE AND WENT OUT OF THE ROOM ON HIS OWN. SINCE THAT TIME THE TABLE DID NOT MOVE IN THE VERTICAL DIRECTION. THIS REPORTED ISSUE OCCURRED IN (B)(6).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SIMVIEW 3000 SIMULATOR SYSTEM, SIMULATION, RADIATION THERAPY KPQ SIEMENS MEDICAL SOLUTIONS USA, INC., RADIATION ONCOLOGY 01957153 NA

Patients

Seq Age Sex Outcome Treatment
1 TABLE