FDA Adverse Event Malfunction Summary report: N

INFANT DUAL-HEATED EVAQUA BREATHING CIRCUIT

MDR report key: 2443124 · Received February 8, 2012

Report

Report Number
9611451-2012-00062
Event Type
Malfunction
Date Received
February 8, 2012
Date of Event
January 16, 2012
Report Date
January 16, 2012
Manufacturer
FISHER & PAYKEL HEALTHCARE LIMITED
Product Code
BZE
PMA / PMN Number
K034026
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). METHOD: ONLY THE PRESSURE LINE OF THE COMPLAINT RT235 INFANT DUAL-HEATED EVAQUA BREATHING CIRCUIT WAS RETURNED TO FPH IN (B)(4) FOR INSPECTION. THE RETURNED COMPONENT WAS VISUALLY INSPECTED FOR MISSING COMPONENTS. RESULTS: VISUAL INSPECTION REVEALED THAT THE NEO PRESSURE ELBOW OF THE PRESSURE LINE WAS MISSING, CONFIRMING THE REPORTED FAULT. A LOT CHECK REVEALED NO OTHER COMPLAINTS OF THIS NATURE FOR LOT NUMBER 110729. CONCLUSION: EACH BREATHING CIRCUIT KIT CONSISTS OF A NUMBER OF COMPONENTS GROUPED TOGETHER DURING PRODUCTION. IT IS LIKELY THAT OPERATOR ERROR HAS RESULTED IN THE OMITTED PRESSURE ELBOW. THERE ARE STANDARD OPERATING PROCEDURES (SOPS) IN PLACE TO ASSIST OPERATORS ON THE PRODUCTION LINE TO CORRECTLY PACK THE BREATHING CIRCUITS. THIS CONSISTS OF A SPECIFIC PACK CARD DETAILING ALL COMPONENTS REQUIRED WHICH MUST BE DISPLAYED AT THE PACKING STATION. (B)(4).

Additional Manufacturer Narrative · 1

(B)(4). THE COMPLAINT RT235 INFANT DUAL-HEATED EVAQUA BREATHING CIRCUIT IS EN ROUTE TO FISHER & PAYKEL HEALTHCARE FOR INVESTIGATION. WE WILL PROVIDE A FOLLOW-UP REPORT ONCE WE RECEIVE THE COMPLAINT DEVICE AND HAVE COMPLETED OUR INVESTIGATION.

Description of Event or Problem · 1

A HOSPITAL IN (B)(6) REPORTED TO A FISHER & PAYKEL HEALTHCARE (FPH) FIELD REPRESENTATIVE THAT THE PRESSURE ELBOW CONNECTOR OF AN RT235 INFANT DUAL-HEATED EVAQUA BREATHING CIRCUIT WAS NOT INCLUDED IN THE KIT.

Description of Event or Problem · 1

A HOSPITAL IN (B)(6) REPORTED TO A FISHER & PAYKEL HEALTHCARE (FPH) FIELD REPRESENTATIVE THAT THE PRESSURE ELBOW CONNECTOR OF AN RT235 INFANT DUAL-HEATED EVAQUA BREATHING CIRCUIT WAS NOT INCLUDED IN THE KIT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 INFANT DUAL-HEATED EVAQUA BREATHING CIRCUIT BZE BZE FISHER & PAYKEL HEALTHCARE LIMITED RT235 110729

Patients

Seq Age Sex Outcome Treatment
1