UDI-DI
10843997011775
PMA / PMN Number
K203113
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

OTHER TEXT: DEVICE EVALUATION IS ANTICIPATED. AT THE CONCLUSION OF OUR INVESTIGATION OR IF NEW INFORMATION IS OBTAINED, A FOLLOW-UP REPORT WILL BE SUBMITTED. E1. INITIAL REPORTER ZIP CODE EXCEEDED THE MAXIMUM ALLOWABLE CHARACTERS; ZIP CODE IS AS FOLLOWS: (B)(6).

Additional Manufacturer Narrative · 0

OTHER, OTHER TEXT: THE RETURNED SENSOR WAS EVALUATED. THE SENSOR WAS FOUND TO HAVE AN OPEN TRACE AT THE TIP OF THE SENSOR CONNECTOR. THE RETURNED SENSOR WAS USED; THEREFORE, ACCURACY TESTING COULD NOT BE PERFORMED. THE REPORTED ISSUE COULD NOT BE REPLICATED DURING TESTING. B3. CORRECTED "EVENT DATE" FROM 01/01/1901 TO 01/01/2026. E1. INITIAL REPORTER ZIP CODE EXCEEDED THE MAXIMUM ALLOWABLE CHARACTERS; ZIP CODE IS AS FOLLOWS: (B)(6).

Description of Event or Problem · 0

THE CUSTOMER REPORTED THAT DURING ANESTHESIA, VALUES INDICATIVE OF AN AWAKE STATE WERE OBSERVED. AFTER REPLACEMENT OF THE DISPOSABLE COMPONENTS, VALUES CONSISTENT WITH ANESTHESIA WERE OBSERVED. THERE WAS NO PATIENT IMPACT OR CONSEQUENCE REPORTED.

Description of Event or Problem · 0

THE CUSTOMER REPORTED THAT DURING ANESTHESIA, VALUES INDICATIVE OF AN AWAKE STATE WERE OBSERVED. AFTER REPLACEMENT OF THE DISPOSABLE COMPONENTS, VALUES CONSISTENT WITH ANESTHESIA WERE OBSERVED. THERE WAS NO PATIENT IMPACT OR CONSEQUENCE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
449295 SEDLINE EEG SENSOR ELECTROENCEPHALOGRAPH OLW MASIMO - 15750 ALTON PKWY 4248 25F03 10843997011775

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown