INSYTE AUTOGUARD
Report
- Report Number
- 1710034-2026-00180
- Event Type
- Malfunction
- Date Received
- February 23, 2026
- Date of Event
- February 10, 2026
- Report Date
- February 27, 2026
- Manufacturer
- BECTON DICKINSON INFUSION THERAPY SYSTEMS INC.
- Product Code
- FOZ
- UDI-DI
- 00382903825370
- PMA / PMN Number
- K201075
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
A DEVICE HISTORY RECORD REVIEW WAS COMPLETED BY OUR QUALITY ENGINEER TEAM FOR PROVIDED MATERIAL NUMBER 382537 AND LOT NUMBER 5161790. THE REVIEW DID NOT REVEAL ANY DETECTED ABNORMALITIES DURING THE PRODUCTION PROCESS THAT COULD HAVE CONTRIBUTED TO THE REPORTED ISSUE, AND ALL QUALITY TESTS WERE FOUND TO BE WITHIN SPECIFICATION. AS A SAMPLE WAS UNAVAILABLE FOR RETURN, A THOROUGH SAMPLE INVESTIGATION COULD NOT BE COMPLETED. BASED ON THE INVESTIGATION RESULTS, AN EXACT CAUSE FOR THIS INCIDENT COULD NOT BE IDENTIFIED. SHOULD YOU AGAIN EXPERIENCE ANY PROBLEMS WITH OUR PRODUCT WE WOULD APPRECIATE THE OPPORTUNITY TO CONDUCT A THOROUGH ANALYSIS. THERE ARE QUALITY CONTROLS CURRENTLY IN PLACE TO DETECT SIMILAR ISSUES DURING THE PRODUCTION PROCESS. FURTHER ACTION HAS NOT BEEN DETERMINED NECESSARY AT THIS TIME. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. OUR QUALITY TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS.
H.3. A FOLLOW UP MDR WILL BE SUBMITTED IF ADDITIONAL INFORMATION, A DEVICE EVALUATION, OR A DEVICE HISTORY REVIEW IS COMPLETED.
NO NEW INFORMATION.
IT WAS REPORTED THAT "I TESTED ANOTHER CATHETER FROM THE SAME LOT AND RETRACTED SLOWLY AS WELL."
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 481253 | INSYTE AUTOGUARD | PERIPHERAL IV CATHETERS | FOZ | BECTON DICKINSON INFUSION THERAPY SYSTEMS INC. | 5161790 | 00382903825370 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |