FDA Adverse Event Malfunction Summary report: N

*

MDR report key: 2441080 · Received January 31, 2012

Report

Report Number
2441080
Event Type
Malfunction
Date Received
January 31, 2012
Date of Event
January 20, 2012
Report Date
January 31, 2012
Manufacturer
BECTON DICKINSON
Product Code
JSD
Product Problem
Yes
Report Source
User Facility report
Reporter Location
IL, US

Narratives

Description of Event or Problem · 1

THE TRYPTICASE SOY BROTH (TSB) WAS USED TO PLATE A CHEST ABSCESS SWAB SPECIMEN AND TO PERFORM A GRAM STAIN. THERE WAS NO GROWTH OF GRAM POSITIVE BACILLI ON THE CULTURE PLATE. THE GRAM STAINED SLIDE WAS PULLED AND REVIEWED. IT WAS DISCOVERED AFTER GRAM STAINING THE TSB ALONE, THAT THE SOURCE OF THE GRAM POSITIVE BACILLI WAS THE BROTH ITSELF. THE GRAM STAIN RESULT WAS CANCELLED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 * TRYPTICASE SOY BROTH JSD BECTON DICKINSON * 1245773

Patients

Seq Age Sex Outcome Treatment
1 76 YR