FDA Adverse Event Injury Summary report: N

SILICONE ULTRAVIOLET-ABSORBING POSTERIOR CHAMBER THREE PIECE FOLDABLE INTRAOCULA

MDR report key: 2440723 · Received January 31, 2012

Report

Report Number
2023826-2012-00070
Event Type
Injury
Date Received
January 31, 2012
Date of Event
January 17, 2012
Report Date
January 20, 2012
Manufacturer
STAAR SURGICAL CO.
Product Code
HQL
PMA / PMN Number
P900048
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). VISUAL INSPECTION OF THE RETURNED PRODUCT FOUND LOOP HAPTIC AND PIECE OF OPTIC STUCK INSIDE CARTRIDGE. NO VISIBLE DAMAGE TO CARTRIDGE. THERE WAS EVIDENCE OF CLEAR SURGICAL RESIDUE. CONCLUSION: AN INVESTIGATION OF THE ROOT CAUSES OF LENS TEARS, SIMILAR TO THAT STATED IN THIS CLAIM, WERE ADDRESSED IN A CAPA OPENED IN JUNE 2005 (WHICH WAS SUBSEQUENTLY CLOSED). THE CAPA ADDRESSED DELIVERY SYSTEM ISSUES INCLUDING ALL STAGES OF MANUFACTURING OF THE INJECTORS AND CARTRIDGES. PROCESSES WERE REVIEWED AND REVISED AS OPPORTUNITIES FOR IMPROVEMENT WERE REVEALED. THE CAPA ALSO ADDRESSED HANDLING ERRORS. ALL INJECTOR/CARTRIDGE DIRECTIONS FOR USE (DFU) HAVE BEEN MODIFIED TO ADD FURTHER CLARIFICATION TO INSTRUCT THE USERS IN THE PROPER DELIVERY TECHNIQUES THAT ARE EFFECTIVE, AND MINIMIZE THE POTENTIAL FOR DAMAGING THE LENS. BASED ON THE COMPLAINT HISTORY AND THE EVALUATION OF THE RETURNED PRODUCT, POSSIBLE ROOT CAUSES FOR LENS TEARS INCLUDE BOTH DELIVERY SYSTEM ISSUES AND HANDLING ERRORS BY THE CUSTOMER. (B)(4).

Description of Event or Problem · 1

THE REPORTER STATED THE SURGEON INSERTED AN AQ2015A SILICONE ASPHERIC THREE PIECE LENS. THE TRAILING HAPTIC BROKE OFF WHILE INSERTING INTO THE EYE. LENS WAS REMOVED, NO ENLARGED INCISION. BACKUP LENS, SAME MODEL AND SIZE WAS IMPLANTED AND ONE SUTURE USED TO CLOSE THE WOUND.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SILICONE ULTRAVIOLET-ABSORBING POSTERIOR CHAMBER THREE PIECE FOLDABLE INTRAOCULA INTRAOCULAR LENS HQL STAAR SURGICAL CO. AQ2015A NA

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention INJECTOR: MODEL MSI-TM, LOT NUMBER UNKNOWN| CARTRIDGE: MODEL AQ CARTRIDGE-FP, LOT NUMBER UNK| VISCOELASTIC: STAARVISC II, LOT NUMBER UNKNOWN