FDA Adverse Event Injury Summary report: N

PUMP 381 PUMP SET (US)

MDR report key: 24404934 · Received February 20, 2026

Report

Report Number
1220648-2026-03696
Event Type
Injury
Date Received
February 20, 2026
Date of Event
October 22, 2025
Manufacturer
ABIOMED, INC. - 1220648
Product Code
OZD
UDI-DI
00813502012279
PMA / PMN Number
P140003
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE IMPELLA DEVICE WAS NOT RECEIVED FROM THE CUSTOMER, THEREFORE, INVESTIGATION OF THE DEVICE WAS NOT POSSIBLE. SHOULD THE DEVICE OR ANY NEW INFORMATION BE RECEIVED, A SUPPLEMENTAL MDR WILL BE FILED.

Additional Manufacturer Narrative · 0

THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY ABIOMED INC, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, ABIOMED INC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE. CORRECTED INFORMATION WAS PROVIDED IN D4. UPON REVIEW, THE PRIMARY UDI NUMBER HAS BEEN UPDATED.

Additional Manufacturer Narrative · 0

CORRECTED: D1, D4 (CATALOG & SERIAL) HEMATURIA: THE CAUSE OF THE INJURY WAS NOT DETERMINED DUE TO INSUFFICIENT CLINICAL DETAILS. ACCESS SITE ADVERSE EVENT: THE CAUSE OF THE ASAE WAS NOT DETERMINED DUE TO INSUFFICIENT CLINICAL DETAILS.

Description of Event or Problem · 0

1M PMH HEART TRANSPLANT IN 2012. PRESENTED IN ACUTE DECOMPENSATED HF W LACTIC OF 2.8, WAS BROUGHT TO CCL FROM CICU FOR R/LHC AND BIOPSY AS PT IS S/P TRANSPLANT. RHC NUMBERS SHOWED VOLUME OVERLOAD AND SLIGHTLY REDUCED CI, IABP PLACED. LHC THEN SHOWED SEVERE LM LESION. PT BROUGHT BACK TO CCL TODAY FOR IMPELLA INSERTION RELATED TO HF SHOCK AND HRPCI. ON ARRIVAL TO CICU, THERE WAS A NEW HEMATOMA AT IMPELLA SITE, SO A FEMO STOP WAS PLACED ABOVE SITE. BLOODY URINE WAS ALSO FOUND. DURING IMPELLA CP IMPLANT PROCEDURE, THERE WERE FLUID LEAK ISSUES DUE TO A DAMAGED 25 CM INTRODUCER. THE ABIOMED REPRESENTATIVE STATED ABNORMAL AMOUNTS OF BLEEDING AROUND THE HEMOSTATIC VALVE AND SHEATH ACCESS SITE. THIS BEGAN AFTER MICROPUNCTURE FOR SINGLE ACCESS. RE-ACCESS OF THE COMPANION SHEATH AT A DIFFERENT ANGLE WAS ATTEMPTED, BUT DID NOT RESOLVE THE ISSUE. THE 25 CM INTRODUCER WAS REMOVED AND EXCHANGED WITH THE 13 CM INTRODUCER, WHICH REQUIRED THE IMPELLA CP TO BE REMOVED AND A DELAY TO TREATMENT/THERAPY. DURING THE REPLACEMENT, THERE WAS AN EVIDENT CRACK ON THE 25 CM INTRODUCER. THE IMPELLA WAS RE-IMPLANTED AND THE 13 CM INTRODUCER WAS UTILIZED AS A COMPANION SHEATH FOR SINGLE ACCESS WITHOUT BLEEDING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
625159 PUMP 381 PUMP SET (US) TEMPORARY NON-ROLLER TYPE LEFT HEART SUPPORT BLOOD PUMP OZD ABIOMED, INC. - 1220648 PUMP 381 PUMP SET (US) 2026756559 00813502012279

Patients

Seq Age Sex Outcome Treatment
1