IMMAGE IMMUNOCHEMISTRY SYSTEM

MDR report key: 2440231 · Received February 6, 2012

Report

Report Number
2050012-2012-00416
Event Type
Malfunction
Date Received
February 6, 2012
Date of Event
January 20, 2012
Report Date
January 20, 2012
Manufacturer
BECKMAN COULTER, INC.
Product Code
JQX
PMA / PMN Number
K962294
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

EVALUATION CODES, RESULTS, CODE 100 (OTHER): LOOSE WASTE TRAP FITTING. (B)(4).

Description of Event or Problem · 1

ON (B)(4) 2012, CUSTOMER REPORTED THAT THE IMMAGE 800 INSTRUMENT GENERATED "VACUUM OUT OF RANGE LOW" ERRORS WHEN THE WASH SOLUTION WAS PRIMED, AND THAT THE PROBES ALSO LEAKED. NO INJURIES WERE REPORTED AND NO ERRONEOUS PATIENT RESULTS WERE GENERATED. FIELD SERVICE ENGINEER (FSE) INSPECTED THE INSTRUMENT AND FOUND A LOOSE WASTE TRAP FITTING. FSE REPAIRED THE FITTING AND NO FURTHER LEAKS WERE REPORTED. REPAIR WAS VERIFIED PER ESTABLISHED PROCEDURES.