FDA Adverse Event Injury Summary report: N

WAVEWRITER ALPHA? PRIME

MDR report key: 24401676 · Received February 20, 2026

Report

Report Number
3006630150-2026-01025
Event Type
Injury
Date Received
February 20, 2026
Date of Event
December 19, 2025
Report Date
April 21, 2026
Manufacturer
BOSTON SCIENTIFIC NEUROMODULATION CORPORATION
Product Code
LGW
UDI-DI
08714729985075
PMA / PMN Number
P030017
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION RESULTS: THE DEVICE WAS NOT RETURNED FOR ANALYSIS; THEREFORE, A TECHNICAL ANALYSIS COULD NOT BE PERFORMED. DEVICE HISTORY RECORD: IT WAS CONFIRMED THIS DEVICE MET MANUFACTURING SPECIFICATION PRIOR TO DISTRIBUTION AND THERE WERE NO MANUFACTURING DEVIATIONS WHICH COULD HAVE CONTRIBUTED TO THE REPORTED EVENT. LABELING REVIEW: REVIEW OF THE INSTRUCTIONS FOR USE (IFU) CONFIRMED THERE WAS RELEVANT CONTENT AND SUFFICIENT GUIDANCE WITH RESPECT TO THE CIRCUMSTANCES DESCRIBED WITHIN THIS COMPLAINT. NO UPDATES ARE REQUIRED TO THE IFU AS A RESULT OF THIS EVENT. INVESTIGATION CONCLUSION: BASED ON ALL AVAILABLE INFORMATION, BOSTON SCIENTIFIC HAS ASSIGNED AN INVESTIGATION CONCLUSION CODE OF KNOWN INHERENT RISK OF DEVICE.

Additional Manufacturer Narrative · 0

ADDITIONAL SUSPECT MEDICAL DEVICE COMPONENT INVOLVED IN THE EVENT: PRODUCT FAMILY: SCS-PADDLE LEADS, UPN: M365SC8336500, MODEL: SC-8336-50, SERIAL: (B)(6), BATCH: 7091432, UDI: (B)(4).

Description of Event or Problem · 0

IT WAS REPORTED THAT THE PATIENT WAS EXPERIENCING INADEQUATE PAIN RELIEF. THE PATIENT UNDERWENT A SPINAL CORD STIMULATOR (SCS) SYSTEM EXPLANT PROCEDURE AND WAS DOING WELL POSTOPERATIVELY. THE EXPLANTED DEVICES WERE DISCARDED BY FACILITY.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE PATIENT WAS EXPERIENCING INADEQUATE PAIN RELIEF. THE PATIENT UNDERWENT A SPINAL CORD STIMULATOR (SCS) SYSTEM EXPLANT PROCEDURE AND WAS DOING WELL POSTOPERATIVELY. THE EXPLANTED DEVICES WERE DISCARDED BY FACILITY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
536795 WAVEWRITER ALPHA? PRIME STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF LGW BOSTON SCIENTIFIC NEUROMODULATION CORPORATION SC-1432 238338 08714729985075

Patients

Seq Age Sex Outcome Treatment
1 80 YR Female Required Intervention