UNI-VENT
MDR report key: 2439707
·
Received January 28, 2012
Report
- Report Number
- 2242630-2005-00029
- Event Type
- Other
- Date Received
- January 28, 2012
- Date of Event
- April 21, 2005
- Report Date
- June 17, 2005
- Manufacturer
- IMPACT INSTRUMENTATION, INC.
- Product Code
- BTL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- OTHER
Narratives
Description of Event or Problem · 1
DATE NOTIFIED: (B)(4) 2005. A MODEL 706 VENTILATOR (S/N (B)(4)) FAILED TO CYCLE. AN INSPECTION OF THE DEVICE REVEALED A DEFECTIVE RESISTOR AND A DEFECTIVE INTEGRATED CIRCUIT. THE RESISTOR AND THE IC WERE REPLACED. THE DEVICE WAS TESTED TO SPECIFICATIONS AND RETURNED TO THE CUSTOMER. NO PT WAS INVOLVED AT THE TIME OF THE FAILURE.