FDA Adverse Event Malfunction Summary report: N

CAPTURE-R READY-ID

MDR report key: 2439674 · Received February 6, 2012

Report

Report Number
1034569-2012-00016
Event Type
Malfunction
Date Received
February 6, 2012
Date of Event
December 22, 2012
Report Date
February 2, 2012
Manufacturer
IMMUCOR, INC.
Product Code
KSZ
PMA / PMN Number
STN 102707
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NM, US
Reporter Occupation
MEDICAL TECHNOLOGIST

Narratives

Additional Manufacturer Narrative · 1

REVIEW OF IMAGES: (B)(4), ID152 21DEC2011 11:44. THE FOLLOWING CELLS ARE JKA POSITIVE: 1,4,5,7,8,9,10,11,12,13 & 14. SAMPLE ID (B)(4). THE ECHO GENERATED NEGATIVE RESULTS FOR ALL CELLS. CELL 1: NEGATIVE VISUALLY APPEARED NEGATIVE. CELL 4: NEGATIVE VISUALLY APPEARED NEGATIVE. CELL 5 NEGATIVE VISUALLY APPEARED NEGATIVE. CELL 7: NEGATIVE VISUALLY APPEARED NEGATIVE. CELL 8: NEGATIVE VISUALLY APPEARED NEGATIVE. CELLS 9-14: NEGATIVE VISUALLY APPEARED POSITIVE. CELLS 9-14 ARE HOMOZYGOUS. UNABLE TO RULE OUT ANTIBODY SHOWING DOSAGE. POS CTRL: 4+ NEG CTRL: (B)(4), EXTEND I DP052 21DEC2011 14:27. THE FOLLOWING CELLS ARE JKA POSITIVE: 1,3,4,5,7,8,9,12 & 14. SAMPLE ID (B)(4). CELL 1: NEGATIVE VISUALLY APPEARS NEGATIVE. CELL 3: NEGATIVE VISUALLY APPEARS NEGATIVE. CELL 4: NEGATIVE VISUALLY APPEARS NEGATIVE. CELL 5: 2+ VISUALLY APPEARS POSITIVE. *CELL 7: NEGATIVE VISUALLY APPEARS POSITIVE. *CELL 8: NEGATIVE VISUALLY APPEARS POSITIVE. *CELL 9: NEGATIVE VISUALLY APPEARS POSITIVE. CELL 12: 1+ VISUALLY APPEARS POSITIVE. *CELL 14: NEGATIVE VISUALLY APPEARS POSITIVE. POS CTRL: 4+ NEG CTRL: 0. (B)(4), 21DEC2011 11:27 CELL 3 IS JKA POSITIVE. CELL 3: NEGATIVE VISUALLY APPEARS POSITIVE. POS CTRL: 4+. NEGATIVE WELL INTERPRETATIONS THAT VISUALLY APPEAR POSITIVE ARE ADDRESSED INTECHNICAL COMMUNICATION (B)(4) ISSUED ON (B)(4) 2009.

Description of Event or Problem · 1

A CUSTOMER REPORTED UNEXPECTED NEGATIVE REACTIONS ON CAPTURE-R READY SCREEN (CRRS) AND CAPTURE-R READY ID (CRRID) WHEN TESTING A PATIENT SAMPLE KNOWN TO HAVE AN ANTI-JKA ON THE ECHO.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPTURE-R READY-ID REAGENT RED BLOOD CELLS KSZ IMMUCOR, INC. ID152

Patients

Seq Age Sex Outcome Treatment
1 78 YR