FDA Adverse Event Malfunction Summary report: N

ALINITY I ANTI-HBS REAGENT KIT

MDR report key: 24389485 · Received February 19, 2026

Report

Report Number
3008344661-2026-00038
Event Type
Malfunction
Date Received
February 19, 2026
Date of Event
February 10, 2026
Report Date
April 29, 2026
Manufacturer
ABBOTT IRELAND DIAGNOSTICS DIVISION
Product Code
LOM
UDI-DI
00380740159900
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

AN EVALUATION IS IN PROCESS. A FINAL REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE. UPDATED SECTION B6: DESCRIBE EVENT OR PROBLEM: ADDITIONAL INFORMATION PROVIDED ON 27FEB2026: SID (B)(6) =HBV DNA TEST RESULT WAS NEGATIVE.

Additional Manufacturer Narrative · 0

AN EVALUATION IS IN PROCESS. A FINAL REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE.SECTION A PATIENT INFORMATION: REFER TO SECTION B5 FOR COMPLETE PATIENT INFORMATION. THIS REPORT IS BEING FILED ON AN INTERNATIONAL PRODUCT, LIST NUMBER 7P89 THAT HAS A SIMILAR PRODUCT DISTRIBUTED IN THE US, LIST NUMBER 7P88, WITH 510K/PMA/BLA NUMBER P050051.

Description of Event or Problem · 0

THE CUSTOMER OBSERVED FALSE REACTIVE ALINITY I ANTI-HBS RESULTS ON FOUR PATIENTS, THAT WERE REPORTED TO MEDICAL PROVIDER. THE FOLLOWING DATA WAS PROVIDED. SID (B)(6), (B)(6) FEMALE = 10.78 MIU/ML, SID (B)(6), 86 YEARS OLD FEMALE =13.23 MIU/ML, SID (B)(6), 48 YEARS OLD FEMALE = 11.57 MIU/ML, SID (B)(6), 52 YEARS OLD MALE =34.86 MIU/ML, THE CUSTOMER USES REFERENCE RANGE =<10 MIU/ML. ADDITIONAL INFORMATION PROVIDED ON 27FEB2026: SID (B)(6) = HBV DNA TEST RESULT WAS NEGATIVE, NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

Description of Event or Problem · 0

THE CUSTOMER OBSERVED FALSE REACTIVE ALINITY I ANTI-HBS RESULTS ON FOUR PATIENTS, THAT WERE REPORTED TO MEDICAL PROVIDER. THE FOLLOWING DATA WAS PROVIDED. SID (B)(6) , 91 YEARS OLD FEMALE = 10.78 MIU/ML SID (B)(6) , 86 YEADS OLD FEMALE =13.23 MIU/ML SID (B)(6) , 48 YEARS OLD FEMALE = 11.57 MIU/ML SID (B)(6) , 52 YEARS OLD MALE =34.86 MIU/ML THE CUSTOMER USES REFERENCE RANGE =<10 MIU/ML NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
447266 ALINITY I ANTI-HBS REAGENT KIT TEST, HEPATITIS B (B CORE, BE ANTIGEN, BE ANTIBODY, B CORE IGM) LOM ABBOTT IRELAND DIAGNOSTICS DIVISION 73442FZ01 00380740159900

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown ALNTY I PROCESSING MODU, 03R65-01, (B)(6).| ALNTY I PROCESSING MODU, 03R65-01, (B)(6).