POWERED WHEELCHAIR
Report
- Report Number
- 3004493922-2012-00035
- Event Type
- Injury
- Date Received
- February 3, 2012
- Date of Event
- January 13, 2012
- Report Date
- February 1, 2012
- Manufacturer
- INVACARE SUZHOU
- Product Code
- ITI
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MO, US
- Reporter Occupation
- PATIENT FAMILY MEMBER OR FRIEND
Narratives
NO RMA HAS BEEN INITIATED FOR THIS ISSUE. MODEL M41SRB, (B)(4) IS APPROXIMATELY 4 YEARS 4 MONTHS OLD. THE USER MANUAL PART NUMBER 1143206 WAS ISSUED WITH THIS DEVICE. THE USER MANUAL IS ALSO FOUND ON-LINE AT INVACARE.COM. IT IS UNKNOWN IF THE CONSUMER HAS FULLY READ AND UNDERSTANDS THE USER MANUAL. DOCUMENTATION PROVIDES WARNINGS, CAUTIONS, AND INSTRUCTIONS FOR SAFELY USING THE DEVICE. IF THE CONSUMER DOES NOT UNDERSTAND THE WRITTEN WARNINGS, CAUTIONS OR INSTRUCTIONS THEN THEY SHOULD CONTACT INVACARE. THE CONSUMER IS A DOUBLE AMPUTEE DUE TO A PREEXISTING MEDICAL CONDITION. THE CONSUMERS STABILITY AND MEDICATION REGIMEN ARE UNKNOWN. THE CONSUMER IS A (B)(6) MALE WHOSE HEIGHT AND WEIGHT ARE UNKNOWN. CONSUMER STATED THAT HE WAS INDOORS IN HIS BED ROOM WHEN THE ALLEGED INCIDENT OCCURRED. HE REACHED DOWN (FORWARD) TO PICK SOMETHING UP WHEN THE SEAT POSITIONING STRAP ALLEGEDLY GAVE WAY CAUSING HIM TO FALL FORWARD ON HIS FACE. HE WENT TO THE EMERGENCY ROOM WHERE THEY DID X-RAYS AND DISCOVERED THAT HE ALLEGEDLY BROKE HIS NECK. THE CONSUMER STATED THAT HE IS CURRENTLY IN A BRACE. INVACARE CONTACTED THE DEALER AND ADVISE THEM THAT WE WOULD LIKE TO HAVE A TECHNICIAN GO OUT TO ASSESS THE UNIT AND REPLACE THE SEAT POSITIONING STRAP AS THE CONSUMER IS STILL USING THE DEVICE. THE CONSUMERS TECHNIQUE WHILE USING THE DEVICE IS UNKNOWN. PAGE 23 OF THE OWNER'S MANUAL STATES A WARNING, "DO NOT ATTEMPT TO REACH OBJECTS IF YOU HAVE TO MOVE FORWARD IN THE SEAT OR PICK THEM UP FROM THE FLOOR BY REACHING DOWN BETWEEN YOUR KNEES". PAGE 40 OF THE OWNER'S MANUAL STATES A WARNING FOR THE SEAT POSITIONING STRAP, "THE SEAT POSITIONING STRAP IS A POSITIONING BELT ONLY. IT IS NOT DESIGNED FOR USE AS A SAFETY DEVISE WITHSTANDING HIGH STRESS LOADS", "IF SIGNS OF WEAR APPEAR, STRAP MUST BE REPLACED IMMEDIATELY". THE MAINTENANCE HISTORY OF THE DEVICE SHOWS THAT THE POWER CHAIR HAD A COMPLETE REFURBISHMENT DONE IN (B)(4) 2011. DEALER STATED THAT THE CONSUMER CALLED THE DEALERSHIP ON (B)(6) 2012 TO HAVE HIS BATTERIES REPLACED AND ADDITIONAL WORK DONE ON THE POWER CHAIR.
USER WAS REACHING DOWN TO PICK SOMETHING UP WHEN THE SEAT POSITIONING STRAP ALLEGEDLY BROKE CAUSING THE USER TO FALL FROM THE CHAIR. EMERGENCY ROOM VISIT & X-RAYS ALLEGEDLY DETERMINED THAT USER BROKE HIS NECK.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | POWERED WHEELCHAIR | 890.3860 | ITI | INVACARE SUZHOU | M41SRB |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 70 | Other |