FDA Adverse Event Malfunction Summary report: N

RESPIRONICS

MDR report key: 24366625 · Received February 17, 2026

Report

Report Number
2518422-2026-101897
Event Type
Malfunction
Date Received
February 17, 2026
Date of Event
January 29, 2026
Report Date
March 2, 2026
Manufacturer
RESPIRONICS, INC.
Product Code
MNT
PMA / PMN Number
K102985
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
BIOMEDICAL ENGINEER
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

PER GOOD FAITH EFFORT (GFE) RESPONSE, IT WAS CONFIRMED THAT THE CUSTOMER RECEIVED A THIRD POWER CORD AND INSTALLED IT. GROUNDS TESTING WAS SUCCESSFUL; THE DEVICE PASSED THE ENTIRE PERFORMANCE VERIFICATION TESTING AND HAS BEEN RETURNED TO SERVICE. THE INVESTIGATION CONCLUDES THAT NO FURTHER ACTION IS REQUIRED AT THIS TIME. IF ADDITIONAL INFORMATION IS RECEIVED THE COMPLAINT FILE WILL BE REOPENED.

Description of Event or Problem · 0

PHILIPS RECEIVED A COMPLAINT FROM A BIOMEDICAL TECHNICIAN REGARDING A V60 VENTILATOR, NOTING WHILE PERFORMING PREVENTATIVE MAINTENANCE (PM) PERFORMANCE VERIFICATION TESTING (PVT), THE DEVICE FAILED THE ELECTRICAL SAFETY TEST, MEASURING 600 MILLIOHMS OF RESISTANCE. A REPLACEMENT POWER CORD WAS ORDERED AND INSTALLED; HOWEVER, IT WAS IDENTIFIED AS AN OUT-OF-BOX FAILURE (DEFOA) BY PHILIPS TECHNICAL SUPPORT. POINT-TO-POINT GROUND TESTING USING A CALIBRATED FLUKE METER CONFIRMED THE NEW POWER CORD WAS FAULTY. A SECOND WARRANTY REPLACEMENT POWER CORD WAS PROVIDED, BUT IT ALSO FAILED OUT OF THE BOX, WITH TESTING TODAY SHOWING A RESISTANCE OF OVER 100 MILI OHMS, AND WHEN CONNECTED TO THE DEVICE, THE GROUNDS TESTING WILL NOT PASS. THE CUSTOMER IS STILL OVER 200 MILI OHMS ON ALL TEST POINTS LISTED IN THE MANUAL. THERE WAS NO PATIENT INVOLVEMENT. DUE TO THE PENDING REPAIR OF THE POWER CORD, THE DEVICE HAS NOT YET BEEN TESTED FOLLOWING THE REPLACEMENT OF THE PRESSURE TRANSDUCER MODULE. THE INVESTIGATION IS ONGOING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
428821 RESPIRONICS VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE MNT RESPIRONICS, INC. V60 V60PLUS VENTILATOR

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown