FDA Adverse Event
Injury
Summary report: N
TANDEM
MDR report key: 2436624
·
Received February 2, 2012
Report
- Report Number
- 1020279-2012-00058
- Event Type
- Injury
- Date Received
- February 2, 2012
- Date of Event
- January 5, 2012
- Report Date
- January 5, 2012
- Manufacturer
- SMITH & NEPHEW, INC.
- Product Code
- JDH
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AL, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE TOP LINER RING HAD SIGNS OF PLASTIC DEFORMATION ON THE INNER RIM WAS LIKELY DUE TO FEMORAL HEAD/NECK IMPINGEMENT. THIS IMPINGEMENT WHEN COUPLED WITH LEVER-OUT FORCES, WHICH MAY HAVE BEEN GENERATED DURING THE REPORTED DISLOCATION, LIKELY RESULTED IN THE DISASSOCIATION OF THE TOP LINER RING AND THE INNER FEMORAL HEAD.
Description of Event or Problem · 1
IT WAS REPORTED THAT REVISION SURGERY WAS PERFORMED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TANDEM | BIPOLAR | JDH | SMITH & NEPHEW, INC. | 04JM13160 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 82 YR | Hospitalization| R |