630G INSULIN PUMP MMT-1715KR 630G
Report
- Report Number
- 2032227-2026-128555
- Event Type
- Malfunction
- Date Received
- February 17, 2026
- Date of Event
- January 27, 2026
- Report Date
- April 27, 2026
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OZO
- UDI-DI
- 000000763000367060
- PMA / PMN Number
- P150001
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- 003
Narratives
PUMP PASSED THE DISPLACEMENT TEST, SELF-TEST, REWIND TEST, PRIME/SEATING TEST, BASIC OCCLUSION TEST, OCCLUSION TEST AND FORCE SENSOR TEST. NO UNEXPECTED NO DELIVERY ALARMS, PRIME/FILL ANOMALIES WERE NOTED DURING TESTING. THUS, AND SOFTWARE WAS UTILIZED AND DOWNLOADED TRACE/HISTORY FILES PROPERLY. ALARMED NO DELIVERY SHOW IN THE TRACE/HISTORY FILES ON (B)(6) 2026 09:40:00.000, (B)(6) 2026 11:19:00.000, (B)(6) 2026 11:33:18.000, (B)(6) 2026 11:38:23.000 DURING BOLUS. NO MOISTURE DAMAGE ON THE ELECTRONICS, BATTERY CONNECTOR, BATTERY TUBE, MOTOR, VIBRATOR DURING VISUAL INSPECTION. TEST P-CAP AND RESERVOIR LOCKED PROPERLY INTO RESERVOIR COMPARTMENT DURING TESTING. THE FOLLOWING WERE NOTED DURING VISUAL INSPECTION: UNIT HAD SCRATCHED CASE, STAINED KEYPAD OVERLAY. IN CONCLUSION, PUMP PASSED ALL TESTING REQUIRED. INSULIN FLOW BLOCKED/NO DELIVERY ALARMS NOT CONFIRMED. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.
THIS MDR RELATED TO THE PUERTO RICO MANUFACTURING SITE HAS BEEN ASSIGNED A MEDWATCH NUMBER FROM THE MEDTRONIC MINIMED NORTHRIDGE SITE, PER VARIANCE 5. CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. WE THEREFORE CONSIDER THIS REPORT COMPLETE TO THE BEST OF OUR KNOWLEDGE. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.
IT WAS REPORTED TO MEDTRONIC MINIMED THAT THE CUSTOMER EXPERIENCED AN INSULIN FLOW BLOCKED ALARM. THE CUSTOMER REPORTED NO ADVERSE EVENT. THE EVENT INVOLVED PRODUCT(S) MMT-399A, MMT-332A, MMT-1715KR. TROUBLESHOOTING WAS INITIATED FOR THE INSULIN FLOW BLOCKED ALARM, AND IT WAS IDENTIFIED THAT THE ISSUE WAS A PAST EVENT. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. NO PRODUCT RETURN IS REQUIRED FOR MMT-399A, MMT-332A, MMT-1715KR.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 106659 | 630G INSULIN PUMP MMT-1715KR 630G | AUTOMATED INSULIN DOSING, THRESHOLD SUSPEND | OZO | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-1715KR | HG5UX72 | 000000763000367060 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |