FDA Adverse Event Malfunction Summary report: N

LANX ANTERIOR CERVICAL PLATE SYSTEM

MDR report key: 2435314 · Received January 18, 2012

Report

Report Number
3004485144-2012-00003
Event Type
Malfunction
Date Received
January 18, 2012
Date of Event
October 5, 2011
Report Date
January 6, 2011
Manufacturer
LANX, INC
Product Code
KWQ
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

NO PT INJURY REPORTED IN THIS SPECIFIC EVENT. EVENT CONSIDERED MDR REPORTABLE AS PT INJURY IS POSSIBLE IF EVENT WERE TO RECUR. NO REVISION SURGERY PERFORMED. FEEDBACK FROM SURGEON INDICATED THAT SURGERY IS NOT RECOMMENDED AT THIS TIME AS THE PT HAS NOT SHOWN ANY UNUSUAL SYMPTOMS OR PAIN. DISCUSSION WITH SALES REP ON (B)(4) 2012 INDICATED THAT EVENT MAY BE EXPLAINED DUE TO LIMITED SCREW PURCHASE WITHIN THE BONE DUE TO POOR BONE QUALITY; HOWEVER, THE CAUSE REMAINS UNK. THE IFU INDICATES POOR BONE QUALITY AS A CONTRAINDICATION AS WELL AS LOOSENING OR MOVEMENT/MIGRATION AS A POSSIBLE COMPLICATION.

Description of Event or Problem · 1

AN ANTERIOR CERVICAL PLATE BACKED AWAY FROM THE VERTEBRAE. REVIEW OF X-RAY FILMS INDICATED THAT THE SELF-DRILLING SCREWS AND LOCKING CAPS APPEARED SECURELY IN PLACE WITH THE CERVICAL PLATE.