FDA Adverse Event Other Summary report: N

SYSMEX HEMATOLOGY SYSTEM ACCESSORIES

MDR report key: 2434653 · Received January 30, 2012

Report

Report Number
1422681-2012-00001
Event Type
Other
Date Received
January 30, 2012
Date of Event
December 29, 2011
Report Date
January 3, 2012
Manufacturer
SYSMEX CORPORATION
Product Code
GKZ
PMA / PMN Number
K920544
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
MS, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

METHANOL IS NOT MANUFACTURED OR DISTRIBUTED BY SYSMEX, BUT IS USED TO CLEAN THE CASSETTES USED ON THE SP-1000I. THE MSDS FOR METHANOL INDICATES THE POTENTIAL TOXIC HEALTH EFFECTS OF INHALATION AND DESCRIBES THE METHODS OF PERSONAL PROTECTION AND EXPOSURE CONTROL TO PREVENT EXPOSURE TO THESE HEALTH HAZARDS. DURING F/U ON (B)(4) 2012, WITH THE LABORATORY MGR, SHE REPORTED THAT THE USER INVOLVED IN THE INCIDENT WAS HOSPITALIZED FOR ONE DAY, DID NOT SUFFER ANY LINGERING AFFECTS AND IS BACK TO WORK. ALTHOUGH THE LAB MGR DID NOT DESCRIBE THE METHOD IN WHICH THE USER WAS EXPOSED TO THE METHANOL, SHE REPORTED THEY HAVE SINCE CHANGED THEIR POLICY FOR CLEANING CASSETTES TO INCLUDE USING A FACE SHIELD AND BUTYL GLOVES, AND TO USE A SINK IN A VENTILATED HOOD. NO SIMILAR INCIDENTS HAVE OCCURRED SINCE THIS CHANGE TO THEIR PROCEDURE.

Description of Event or Problem · 1

THE LAB MANAGER REPORTED A USER OF A SYSMEX SP-1000I (SLIDE PREPARER-STAINER) COMPONENT OF THE SYSMEX XE-5000 ALPHA-N BECAME LIGHT-HEADED WHILE PERFORMING THE ROUTINE DAILY CLEANING OF THE SLIDE CASSETTES WITH METHANOL. THE USER FELT FAINT AND REPORTED DIFFICULTY WITH READING, SO HE STOPPED WORKING WITH THE CASSETTES AND HAD A COWORKER RELIEVE HIM. HE STATED THE INCIDENT OCCURRED AT 01:30 AM, AND REPORTED IT TO THE LAB MANAGER AT 05:30 AM. THE USER WENT TO THE EMERGENCY DEPT AT 07:00 AM, AND WAS HOSPITALIZED FOR ONE DAY AND RELEASED. NO LINGERING AFFECTS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SYSMEX HEMATOLOGY SYSTEM ACCESSORIES XE-5000 ALPHA-N GKZ SYSMEX CORPORATION SP-1000I

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization