FDA Adverse Event Injury Summary report: N

VACUETTE SAFETY BLOOD COLLECTION SET WITH LUER

MDR report key: 2434447 · Received January 26, 2012

Report

Report Number
8020040-2012-00001
Event Type
Injury
Date Received
January 26, 2012
Date of Event
November 25, 2011
Report Date
January 26, 2012
Manufacturer
GREINER BIO-ONE GMBH
Product Code
FMI
PMA / PMN Number
K011786
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

SBC DEVICE WAS USED AS INFUSION SET FOR 3 DAYS ATTACHED TO PATIENT'S LEFT THIGH SUBCUTANEOUSLY. SBC SET WAS ATTACHED WITH CLEAR PATCH. PATIENT WAS GIVEN MORPHINE VIA A PORT ATTACHED TO THE SBC SET BY THE HOSPITAL. AFTER 3 DAYS OF USE NURSE FOUND SOME BLOOD IN THE TUBING. THEN CHECKED DEVICE BY CAREFULLY REMOVING PATCH AND SAW THAT NEEDLE WAS NOT ATTACHED ANYMORE. X-RAYS LOCATED THE NEEDLE IN PATIENT LEFT THIGH AND HAS MIGRATED TO THE BONE (FEMUR). PATIENT HAD NO SYMPTOMS (E.G. NO PAIN, NO REDNESS, NO FEVER). NEEDLE WAS NOT REMOVED BECAUSE OF PATIENTS CONDITIONS. INITIAL INVESTIGATION SHOWS NO MALFUNCTION OF DEVICE. ACCORDING TO CURRENT INVESTIGATION THE NEEDLE BROKE OFF BY EXTERNAL FORCE DUE TO MOVING OF PATIENT OVER 3 DAYS. ADDITIONALLY DEVICE WAS ATTACHED UNDER BLANKET FOR 3 DAYS AND WAS NOT CLEAR VISIBLE WHEN PATIENT WAS MOVED OR WHEN PATIENT MOVED.

Description of Event or Problem · 1

SBC DEVICE WAS USED AS INFUSION SET FOR 3 DAYS ATTACHED TO PATIENT'S LEFT THIGH SUBCUTANEOUSLY. SBC SET WAS ATTACHED WITH CLEAR PATCH. PATIENT WAS GIVEN MORPHINE VIA A PORT ATTACHED TO THE SBC SET BY THE HOSPITAL. AFTER 3 DAYS OF USE NURSE FOUND SOME BLOOD IN THE TUBING. THE CHECKED DEVICE BY CAREFULLY REMOVING PATCH AND SAW THAT NEEDLE WAS NOT ATTACHED ANYMORE. X-RAYS LOCATED THE NEEDLE IN PATIENT LEFT THIGH AND HAS MIGRATED TO THE BONE (FEMUR). PATIENT HAD NO SYMPTOMS (E.G. NO PAIN, NO REDNESS, NO FEVER). NEEDLE WAS NOT REMOVED BECAUSE OF PATIENT'S CONDITIONS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VACUETTE SAFETY BLOOD COLLECTION SET WITH LUER SBC SET FMI GREINER BIO-ONE GMBH 450099 UNKNOWN

Patients

Seq Age Sex Outcome Treatment
1