LOW PROFILE SCREW, 1.4X12MM, CORTICAL,TI
Report
- Report Number
- 1220246-2026-00599
- Event Type
- Malfunction
- Date Received
- February 16, 2026
- Date of Event
- January 28, 2026
- Report Date
- May 13, 2026
- Manufacturer
- ARTHREX, INC.
- Product Code
- HRS
- UDI-DI
- 00888867303829
- PMA / PMN Number
- K191326
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- DE, US
- Reporter Occupation
- 501
Narratives
INVESTIGATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE PROVIDED UPON AVAILABILITY OF ADDITIONAL INFORMATION.
ON 28-JAN-2026, A SALES REPRESENTATIVE REPORTED VIA EMAIL THAT AN AR-18714-12 CORTICAL SCREW (1.4 MM × 12 MM) EXPERIENCED A HEAD FRACTURE DURING INSERTION. THE SCREW WAS BEING PLACED INTO A HOLE OF A 10-HOLE 1.4 MM PLATE AND HAD BEEN ADVANCED NEARLY FLUSH TO THE PLATE WHEN THE HEAD SEPARATED FROM THE SHAFT. THE SURGEON REMOVED ONLY THE SCREW HEAD, LEAVING THE SCREW SHAFT IN SITU BECAUSE IT CROSSED THE FRACTURE SITE. THE EVENT OCCURRED DURING A METACARPAL FRACTURE INTERNAL FIXATION PROCEDURE ON (B)(6) 2026. THE ISSUE DID NOT CAUSE PROCEDURAL DELAY, AND NO PATIENT HARM WAS REPORTED. ADDITIONAL INFORMATION WAS RECEIVED ON 03-FEB-2026: THERE WAS NO REPORTED DAMAGE TO THE AR-18714P-07 STRAIGHT PLATE, 1.4 MM, 10-HOLE ASSOCIATED WITH THIS EVENT. THE ONLY ISSUE IDENTIFIED INVOLVED A SINGLE AR-18714-12 CORTICAL SCREW (1.4 MM × 12 MM), AS SHOWN IN THE ATTACHED PHOTO. THE PROCEDURE WAS COMPLETED SUCCESSFULLY, AND THE AFFECTED SCREW REMAINED IMPLANTED. ADDITIONAL SCREWS MEASURING 10 MM, 11 MM, 12 MM, AND 15 MM IN LENGTH WERE ALSO UTILIZED AS PART OF THE CONSTRUCT. NO ADVERSE PATIENT EFFECTS WERE REPORTED IN RELATION TO THIS ISSUE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 11880 | LOW PROFILE SCREW, 1.4X12MM, CORTICAL,TI | BONE FIXATION PLATE | HRS | ARTHREX, INC. | LOW PROFILE SCREW, 1.4X12MM, CORTICAL,TI | UNK | 00888867303829 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |