TUBING SET
Report
- Report Number
- 8010762-2026-0000063
- Event Type
- Malfunction
- Date Received
- February 16, 2026
- Date of Event
- February 7, 2026
- Report Date
- April 21, 2026
- Manufacturer
- MAQUET CARDIOPULMONARY GMBH
- Product Code
- DTZ
- UDI-DI
- 04058863005744
- PMA / PMN Number
- K112360
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
A FOLLOW-UP MEDWATCH WILL BE SUBMITTED WHEN ADDITIONAL INFORMATION BECOMES AVAILABLE. NOTE: THIS EVENT OCCURRED ON THE GERMAN MARKET. IT IS A SIGNIFICANTLY SIMILAR DEVICE TO "HLS SET¿ WHICH IS SOLD IN THE USA UNDER PREMARKET SUBMISSION NUMBER K112360. FOR SECTION D4 ALL AVAILABLE IDENTIFYING INFORMATION IS FOR THE OUT OF THE US DEVICE, SUBJECTED TO THIS REPORT. THEREFORE, NO GUDID INFORMATION EXISTS. FURTHERMORE, UDI INFORMATION (PRIMARY DI NUMBER) PROVIDED IN D4 IS FOR HLS SET WITH CATALOG NUMBER 701069073.
THE EVENT OCCURRED IN GERMANY DURING PATIENT TREATMENT, INVOLVING AN HLS SET. IT WAS REPORTED THAT 3-4 HOURS AFTER STARTING THE ECMO PROCEDURE, IT WAS DISCOVERED, THAT THE INVOLVED HYPOTHERMIA DEVICE (DELTASTREAM HC - MANUFACTURED BY XENIOS) WAS NO LONGER FUNCTIONING. THE WATER PUMP WAS NOT WORKING; THERE WAS NO WATER FLOW. SUBSEQUENTLY, IT WAS FOUND THAT BLOOD WAS LEAKING FROM THE WATER CONNECTIONS OF THE HLS SET OXYGENATOR. BLOOD AND BLOOD CLOTS WERE FOUND IN THE TUBING SYSTEM AND IN THE WATER TANK OF THE HYPOTHERMIA DEVICE. AS A RESULT, BOTH THE HLS SET AND THE HYPOTHERMIA DEVICE WERE REPLACED. THE FURTHER TREATMENT OF THE PATIENT PROCEEDED WITHOUT ANY COMPLICATIONS. NO HARM TO ANY PERSON HAS BEEN REPORTED. AS THE HLS SET WAS LEAKING AND HAS BEEN REPLACED DURING PATIENT TREATMENT A REPORT IS REQUIRED. COMPLAINT ID# (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 560551 | TUBING SET | OXYGENATOR, CARDIOPULMONARY BYPASS | DTZ | MAQUET CARDIOPULMONARY GMBH | HLS SET | 3000505729 | 04058863005744 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 69 YR | Female |