FDA Adverse Event Malfunction Summary report: N

MINIMED 780G US SYSTEM BLE CONNECT 3.0 MG/DL

MDR report key: 24337773 · Received February 13, 2026

Report

Report Number
2032227-2026-126875
Event Type
Malfunction
Date Received
February 13, 2026
Date of Event
February 1, 2026
Report Date
April 22, 2026
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OZP
PMA / PMN Number
P160017
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

THIS MDR RELATED TO THE PUERTO RICO MANUFACTURING SITE HAS BEEN ASSIGNED A MEDWATCH NUMBER FROM THE MEDTRONIC MINIMED NORTHRIDGE SITE, PER VARIANCE 5. CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Additional Manufacturer Narrative · 0

UNIT PASSED THE SELF-TEST, DISPLACEMENT TEST, REWIND TEST, PRIME/SEATING TEST, BASIC OCCLUSION TEST, FORCE SENSOR TEST AND OCCLUSION TEST. NO UNEXPECTED INSULIN FLOW BLOCKED ALARMS NOTED DURING TESTING, AND P-CAP LOCKS PROPERLY INTO THE RESERVOIR COMPARTMENT. UNIT SUCCESSFULLY DOWNLOADED TO THUMP. CONFIRMED PUMP ALARMED INSULIN FLOW BLOCKED ALARM ON 02/01/2026 03:08:20.000 BOLUS, 02/01/2026 04:30:08.000 BOLUS, 02/01/2026 04:37:34.000 BOLUS, 02/01/2026 04:39:47.000 BOLUS AND 02/01/2026 04:50:00.000 BOLUS IN PUMP DOWNLOADED HISTORY DURING BOLUS. PUMP WAS CUT TO PERFORM VISUAL INSPECTION AND FOUND EVIDENCE OF MOISTURE DAMAGE ON PCBA 1 AND PCBA 2. THE FOLLOWING WERE NOTED DURING VISUAL INSPECTION: PILLOWING KEYPAD OVERLAY, STAINED KEYPAD OVERLAY, SCRATCHED CASE, CRACKED CASE-CORNER OF BELT CLIP RAILS AND BATTERY TUBE THREADS - CRACKED. NO DELIVERY/OCCLUSION ALARM, UNEXPECTED INSULIN FLOW BLOCKED ALARM WAS NOT CONFIRMED. COSMETIC DAMAGES WERE NOTED DURING VISUAL INSPECTION. EXPOSED TO MOISTURE CONFIRMED. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Description of Event or Problem · 0

IT WAS REPORTED TO MEDTRONIC MINIMED THAT THE CUSTOMER EXPERIENCED AN INSULIN FLOW BLOCKED ALARM. THE CUSTOMER REPORTED NO ADVERSE EVENT. THE EVENT INVOLVED PRODUCT(S) MMT-1884. TROUBLESHOOTING WAS PERFORMED, AND THE TUBING WAS NOT BENT OR KINKED. THE INSULIN PUMP WILL RETURN AND REPLACE PUMP. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. MMT-1884 WAS REQUESTED AND CUSTOMER RESPONSE WAS THE DEVICE WOULD BE RETURNED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
399815 MINIMED 780G US SYSTEM BLE CONNECT 3.0 MG/DL AUTOMATED INSULIN DOSING DEVICE SYSTEM, SINGLE HORMONAL CONTROL OZP MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-1884 NG3867571H

Patients

Seq Age Sex Outcome Treatment
1 65 YR Male