FDA Adverse Event Malfunction Summary report: N

QUANTUM HEATER-COOLER MAIN UNIT

MDR report key: 24337747 · Received February 13, 2026

Report

Report Number
3006073153-2026-00026
Event Type
Malfunction
Date Received
February 13, 2026
Date of Event
January 15, 2026
Report Date
March 13, 2026
Manufacturer
SPECTRUM MEDICAL LTD
Product Code
DWC
UDI-DI
05060434420961
PMA / PMN Number
K212657
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

ROOT CAUSE DETERMINATION IS PENDING RESULTS OF INVESTIGATION.

Additional Manufacturer Narrative · 0

DURING INVESTIGATION, THE REPORTED FAULT WAS UNABLE TO BE CONFIRMED. IT IS SUSPECTED THAT PRESSURE BUILDUP WAS RELIEVED WHEN CONNECTED TO THE HEATER COOLER CHARGER AS INTENDED. AS PART OF PREVENTATIVE MAINTENANCE, SPECTRUM MEDICAL UPGRADED THE SOLENOID FIRMWARE AND CONFIRMED VIA FUNCTIONAL TESTING THAT THE UNIT WAS ABLE TO COOL TO THE SPECIFIED SETPOINT.

Description of Event or Problem · 0

USER REPORTED THAT THE DEVICE WAS UNABLE TO MEET SETPOINT WHEN TRYING TO COOL. THERE WAS NO PATIENT HARM.

Description of Event or Problem · 0

USER REPORTED THAT THE DEVICE WAS UNABLE TO MEET SETPOINT WHEN TRYING TO COOL. THERE WAS NO PATIENT HARM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
401083 QUANTUM HEATER-COOLER MAIN UNIT TEMPERATURE CONTROLLER DWC SPECTRUM MEDICAL LTD QHC1 05060434420961

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown