FDA Adverse Event Malfunction Summary report: N

ALINITY C MAGNESIUM

MDR report key: 24337539 · Received February 13, 2026

Report

Report Number
3005094123-2026-00071
Event Type
Malfunction
Date Received
February 13, 2026
Date of Event
January 25, 2026
Report Date
February 20, 2026
Manufacturer
ABBOTT IRELAND DIAGNOSTICS DIVISION
Product Code
JGJ
UDI-DI
00380740176525
PMA / PMN Number
K181748
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IT
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE. SECTION A PATIENT INFORMATION: REFER TO SECTION B5 FOR COMPLETE PATIENT INFORMATION.

Additional Manufacturer Narrative · 0

AFTER FURTHER EVALUATION, THE SUSPECT MEDICAL DEVICE WAS CHANGED FROM ALINITY C MAGNESIUM, LIST NUMBER 08P19-25, ABBOTT IRELAND DIAGNOSTICS DIVISION TO ALINITY C PROCESSING MODULE, LIST NUMBER 03R67-01, ABBOTT LABORATORIES. MDR NUMBER 3016438761-2026-00102 HAS BEEN SUBMITTED AND ALL FURTHER INFORMATION WILL BE DOCUMENTED UNDER THAT MDR NUMBER.

Description of Event or Problem · 0

THE CUSTOMER OBSERVED FALSELY ELEVATED MAGNESIUM ON THE ALINITY C PROCESSING MODULE FOR A 46-YEAR-OLD FEMALE. THE SAMPLES WERE REPEATED ON ANOTHER ANALYZER WITH LOWER RESULTS. THE FOLLOWING INFORMATION WAS PROVIDED (REFERENCE RANGE 1.9 - 2.5 MG/DL): SID: (B)(6), INITIAL MAGNESIUM RESULT= 7.66 MG/DL; REPEAT RESULT ON THE OTHER ANALYZER= 1.51 MG/DL. THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.

Description of Event or Problem · 0

THE CUSTOMER OBSERVED FALSELY ELEVATED MAGNESIUM ON THE ALINITY C PROCESSING MODULE FOR A 46-YEAR-OLD FEMALE. THE SAMPLES WERE REPEATED ON ANOTHER ANALYZER WITH LOWER RESULTS. THE FOLLOWING INFORMATION WAS PROVIDED (REFERENCE RANGE 1.9 - 2.5 MG/DL): SID (B)(6), INITIAL MAGNESIUM RESULT= 7.66 MG/DL; REPEAT RESULT ON THE OTHER ANALYZER= 1.51 MG/DL. THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
391530 ALINITY C MAGNESIUM PHOTOMETRIC METHOD, MAGNESIUM JGJ ABBOTT IRELAND DIAGNOSTICS DIVISION UNKNOWN 00380740176525

Patients

Seq Age Sex Outcome Treatment
1 46 YR Female ALNTY C PROCESSING MODU, 03R67-01, (B)(6).| ALNTY C PROCESSING MODU, 03R67-01, (B)(6).