ALINITY C MAGNESIUM
Report
- Report Number
- 3005094123-2026-00071
- Event Type
- Malfunction
- Date Received
- February 13, 2026
- Date of Event
- January 25, 2026
- Report Date
- February 20, 2026
- Manufacturer
- ABBOTT IRELAND DIAGNOSTICS DIVISION
- Product Code
- JGJ
- UDI-DI
- 00380740176525
- PMA / PMN Number
- K181748
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IT
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE. SECTION A PATIENT INFORMATION: REFER TO SECTION B5 FOR COMPLETE PATIENT INFORMATION.
AFTER FURTHER EVALUATION, THE SUSPECT MEDICAL DEVICE WAS CHANGED FROM ALINITY C MAGNESIUM, LIST NUMBER 08P19-25, ABBOTT IRELAND DIAGNOSTICS DIVISION TO ALINITY C PROCESSING MODULE, LIST NUMBER 03R67-01, ABBOTT LABORATORIES. MDR NUMBER 3016438761-2026-00102 HAS BEEN SUBMITTED AND ALL FURTHER INFORMATION WILL BE DOCUMENTED UNDER THAT MDR NUMBER.
THE CUSTOMER OBSERVED FALSELY ELEVATED MAGNESIUM ON THE ALINITY C PROCESSING MODULE FOR A 46-YEAR-OLD FEMALE. THE SAMPLES WERE REPEATED ON ANOTHER ANALYZER WITH LOWER RESULTS. THE FOLLOWING INFORMATION WAS PROVIDED (REFERENCE RANGE 1.9 - 2.5 MG/DL): SID: (B)(6), INITIAL MAGNESIUM RESULT= 7.66 MG/DL; REPEAT RESULT ON THE OTHER ANALYZER= 1.51 MG/DL. THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.
THE CUSTOMER OBSERVED FALSELY ELEVATED MAGNESIUM ON THE ALINITY C PROCESSING MODULE FOR A 46-YEAR-OLD FEMALE. THE SAMPLES WERE REPEATED ON ANOTHER ANALYZER WITH LOWER RESULTS. THE FOLLOWING INFORMATION WAS PROVIDED (REFERENCE RANGE 1.9 - 2.5 MG/DL): SID (B)(6), INITIAL MAGNESIUM RESULT= 7.66 MG/DL; REPEAT RESULT ON THE OTHER ANALYZER= 1.51 MG/DL. THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 391530 | ALINITY C MAGNESIUM | PHOTOMETRIC METHOD, MAGNESIUM | JGJ | ABBOTT IRELAND DIAGNOSTICS DIVISION | UNKNOWN | 00380740176525 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 46 YR | Female | ALNTY C PROCESSING MODU, 03R67-01, (B)(6).| ALNTY C PROCESSING MODU, 03R67-01, (B)(6). |