FDA Adverse Event Malfunction Summary report: N

SMARTDRIVE

MDR report key: 24332627 · Received February 13, 2026

Report

Report Number
3008370857-2026-00320
Event Type
Malfunction
Date Received
February 13, 2026
Date of Event
January 20, 2026
Report Date
March 18, 2026
Manufacturer
MAX MOBILITY LLC
Product Code
ITI
UDI-DI
00861896000310
PMA / PMN Number
K151199
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WA, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

REPORT INDICATES A FAILURE HAVING OCCURRED WHERE THE SMART DRIVE DEVICE REPORTEDLY INITIATED A DRIVE COMMAND WITHOUT PHYSICAL MANIPULATION OF THE SWITCH CONTROLS. REPORT CLAIMS THE END-USER ATTEMPTED TO DISENGAGE A DRIVE COMMAND VIA THE SWITCH CONTROL BUTTONS AND THE DEVICE REPORTEDLY CONTINUED TO DRIVE. THIS REPORTEDLY FORCED THE END-USER TO MANEUVER THE WHEELCHAIR INTO AN IMMOVEABLE OBJECT WHERE THEY SUSTAINED MINOR BRUISING INJURIES TO THEIR LEGS. THE END-USER REPORTS THEY HAVE ONE SWITCH MOUNTED IN THE FRONT OF THE WHEELCHAIR FOR THEIR ACCESS, AND THE SECOND SWITCH MOUNTED TO THE PUSH HANDLE AREA FOR AN ATTENDANT. THE END-USER WAS PROVIDED A REPLACEMENT SET OF SWITCH CONTROLS, WITH THE SUSPECT CONTROLLER BEING RETURNED TO PERMOBIL FOR EVALUATION. AT THIS TIME, PERMOBIL HAS YET TO RECEIVE THE SUSPECT COMPONENT, THUS IS UNABLE TO REACH A DETERMINATION AS TO POSSIBLE CAUSE FOR THE ALLEGED MALFUNCTION. IF ANY NEW INFORMATION IS RECEIVED, A FOLLOW-UP REPORT WILL BE SUBMITTED. A REVIEW OF THE DEVICE HISTORY RECORD FROM THE DATE THE DEVICE WAS MANUFACTURED WAS PERFORMED WHICH NOTED THAT THIS DEVICE WAS NOT INVOLVED IN MANUFACTURING QUALITY NON-CONFORMANCE.

Additional Manufacturer Narrative · 0

PERMOBIL RECEIVED SUSPECT CONTROLLER FOR INSPECTION AND EVALUATION. THROUGH TESTING, PERMOBIL WAS UNABLE TO REPLICATE ANY FAILURE OF INVOLUNTARY ACTIVATION AS ALLEGED, NOR NOTED ANY SIGNS OF A PRODUCT MALFUNCTION HAVING OCCURRED WITH THE SWITCH CONTROLLER, FINDING IT TO REMAIN FULLY OPERATIONAL AS PER DESIGN. PERMOBIL IS UNABLE TO REACH A DETERMINATION AS TO POSSIBLE ROOT CAUSE OF THE ALLEGED MALFUNCTION WITHOUT SPECULATION.

Description of Event or Problem · 0

REPORTS WHILE IN A GROCERY STORE, THE SMARTDRIVE UNIT ENGAGED A DRIVE COMMAND WITHOUT ANY PHYSICAL MANIPULATION OF THE SWITCH CONTROL BUTTON, REPORTEDLY CAUSING THE END-USER TO IMPACT AN IMMOVEABLE OBJECT WITH THE WHEELCHAIR. NO SERIOUS INJURIES WERE REPORTED TO HAVE OCCURRED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
255879 SMARTDRIVE SMARTDRIVE ITI MAX MOBILITY LLC MX2+ 00861896000310

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown