FDA Adverse Event Malfunction Summary report: N

CAPTURE-R READY INDICATOR RED CELLS

MDR report key: 2432931 · Received January 31, 2012

Report

Report Number
1034569-2012-00013
Event Type
Malfunction
Date Received
January 31, 2012
Date of Event
January 4, 2012
Report Date
January 30, 2012
Manufacturer
IMMUCOR, INC.
Product Code
KSZ
PMA / PMN Number
BK020053
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AL, US
Reporter Occupation
MEDICAL TECHNOLOGIST

Narratives

Additional Manufacturer Narrative · 1

A REVIEW OF THE IMAGES FROM BATCH (B)(4) FOR SAMPLE (B)(4) SHOWS THAT CELL 2 IS WEAKLY POSITIVE, BUT CALL NEGATIVE BY THE INSTRUMENT. THE SAMPLE WAS RUN ON 3_CELL SCREEN LOT R192 USING CAPTURE-R READY INDICATOR RED CELLS (CRRIC) LOT 221751. THE SAMPLE WAS RE-TESTED ON (B)(6) 2012, USING 3_CELL SCREEN LOT R192 AND CRRIC LOT 221755. A REVIEW OF THE IMAGES FROM BATCH (B)(4) FOR SAMPLE (B)(4) SHOWS THAT CELL 2 IS 3+ POSITIVE. THE CELL LOOKS POSITIVE. A SECOND SAMPLE, ID (B)(4) WAS DRAWN AND TESTED ON (B)(6) 2012, ON BATCH (B)(4). A REVIEW OF THE IMAGES FROM BATCH (B)(4) FOR SAMPLE (B)(4) SHOWS THAT CELL 2 IS 4+ POSITIVE. THE CELL LOOKS POSITIVE. SWITCHING TO A NEW LOT OF CRRIC RESOLVED THE ISSUE. CRRIC LOT 221751 EXPIRED PRIOR TO INVESTIGATION. REVIEW OF RECORDS SHOWS RETENTION TESTING PERFORMED WHILE PRODUCT WAS IN DATE RESULTED AS EXPECTED.

Description of Event or Problem · 1

A CUSTOMER REPORTED AN UNEXPECTED NEGATIVE ON THE 3_CELL SCREEN ON THE ECHO.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPTURE-R READY INDICATOR RED CELLS ANTI-IGG-COATED INDICATOR RED CELLS FOR USE IN SOLID PHASE ASSAYS KSZ IMMUCOR, INC. 221751

Patients

Seq Age Sex Outcome Treatment
1 48 MO