CAPTURE-R READY INDICATOR RED CELLS
Report
- Report Number
- 1034569-2012-00013
- Event Type
- Malfunction
- Date Received
- January 31, 2012
- Date of Event
- January 4, 2012
- Report Date
- January 30, 2012
- Manufacturer
- IMMUCOR, INC.
- Product Code
- KSZ
- PMA / PMN Number
- BK020053
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AL, US
- Reporter Occupation
- MEDICAL TECHNOLOGIST
Narratives
A REVIEW OF THE IMAGES FROM BATCH (B)(4) FOR SAMPLE (B)(4) SHOWS THAT CELL 2 IS WEAKLY POSITIVE, BUT CALL NEGATIVE BY THE INSTRUMENT. THE SAMPLE WAS RUN ON 3_CELL SCREEN LOT R192 USING CAPTURE-R READY INDICATOR RED CELLS (CRRIC) LOT 221751. THE SAMPLE WAS RE-TESTED ON (B)(6) 2012, USING 3_CELL SCREEN LOT R192 AND CRRIC LOT 221755. A REVIEW OF THE IMAGES FROM BATCH (B)(4) FOR SAMPLE (B)(4) SHOWS THAT CELL 2 IS 3+ POSITIVE. THE CELL LOOKS POSITIVE. A SECOND SAMPLE, ID (B)(4) WAS DRAWN AND TESTED ON (B)(6) 2012, ON BATCH (B)(4). A REVIEW OF THE IMAGES FROM BATCH (B)(4) FOR SAMPLE (B)(4) SHOWS THAT CELL 2 IS 4+ POSITIVE. THE CELL LOOKS POSITIVE. SWITCHING TO A NEW LOT OF CRRIC RESOLVED THE ISSUE. CRRIC LOT 221751 EXPIRED PRIOR TO INVESTIGATION. REVIEW OF RECORDS SHOWS RETENTION TESTING PERFORMED WHILE PRODUCT WAS IN DATE RESULTED AS EXPECTED.
A CUSTOMER REPORTED AN UNEXPECTED NEGATIVE ON THE 3_CELL SCREEN ON THE ECHO.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CAPTURE-R READY INDICATOR RED CELLS | ANTI-IGG-COATED INDICATOR RED CELLS FOR USE IN SOLID PHASE ASSAYS | KSZ | IMMUCOR, INC. | 221751 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 48 MO |