CW HIGH RES SSP UNITRAYS KIT WITH TAQ
Report
- Report Number
- 2244574-2012-00001
- Event Type
- Malfunction
- Date Received
- January 11, 2012
- Date of Event
- October 3, 2011
- Report Date
- January 10, 2012
- Manufacturer
- LIFE TECHNOLOGIES CORPORATION
- Product Code
- MZI
- PMA / PMN Number
- BK020068
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AR, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
USING A SURVEY SAMPLE, THE CUSTOMER REPORTED THAT THERE WAS A FALSE POSITIVE BAND SEEN IN LANE 89 FOR A CONFIRMED SAMPLE TYPE OF HLA C 02:14. LANE 89 WAS LISTED ON THE KIT WORKSHEET/INTERPRETATION DOCUMENTATION (WHICH IS INCLUDED WITH EACH KIT) AS THE FOLLOWING: C 02:10^, HOWEVER THE DOCUMENTATION ALSO INDICATED THAT "ALLELES WRITTEN WITH THE "^" SYMBOL FOLLOWING THEM HAVE UNK SEQUENCES AT THE SENSE AND/OR ANTISENSE PRIMER ANNEALING SITE. THEIR REACTION PATTERNS ARE PREDICTED BASED ON SEQUENCE HEMOLOGY WITH EACH MAJOR HLA GROUP. INTERNAL INVESTIGATION CONFIRMED BY SEQUENCE BASED TYPING THAT THE CUSTOMER HAD A C 02:14 ALLELE. THEREFORE LANE 89 HAS A CONFIRMED FALSE POSITIVE IN THE AFFECTED KITS. THE C 02:14 ALLELE HAS AN UNK SEQUENCE REGION WHICH HAD NOT BEEN CONFIRMED THROUGH TESTING. IT IS NOW KNOWN THAT THE C 02:14 ALLELE HAS THE SAME POLYMORPHISM AS C 02:10 AND THE WORKSHEET IS INCORRECT IN LISTING THIS ALLELE AS NEGATIVE FOR LANE 89. WHEN AN ALLELE, SUCH AS THE C 02:14 ALLELE, CONTAINS A REGION WITH UNK DNA SEQUENCE, IT IS ASSIGNED ACCORDING TO THE COMMON ALLELES IN THAT MOLECULAR HLA GROUP. NO C02:14 SAMPLE WAS AVAILABLE FOR TESTING PRIOR TO THE COMPLAINT; THE LABELING STATED THE PRIMERS IN LANE 89 WOULD NOT AMPLIFY THIS SEQUENCE IN THE C 02:14 ALLELE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CW HIGH RES SSP UNITRAYS KIT WITH TAQ | MZI TEST, QUALITATIVE, FOR HLA, NON-DIAGNOSTIC | MZI | LIFE TECHNOLOGIES CORPORATION | 005 928442 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |