OLYMPUS SINGLE USE MECHANICAL LITHOTRIPTOR V
Report
- Report Number
- 8010047-2012-00012
- Event Type
- Malfunction
- Date Received
- January 19, 2012
- Date of Event
- December 19, 2011
- Report Date
- December 21, 2011
- Manufacturer
- OLYMPUS MEDICAL SYSTEMS CORPORATION
- Product Code
- LQR
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
THE DEVICE REFERENCED IN THIS REPORT WAS NOT RETURNED TO OLYMPUS FOR EVALUATION. THE USER FACILITY REPORTEDLY HAD DISCARDED THE DEVICE FOLLOWING THE PROCEDURE. REVIEW OF COMPLAINT FILE FOR A PERIOD OF 2 YEARS DID NOT IDENTIFY ANY OTHER COMPLAINTS ASSOCIATED WITH THE LOT NUMBER. THE EXACT CAUSE OF THE REPORTED PHENOMENON COULD NOT BE CONCLUSIVELY DETERMINED BUT THE SIZE OR HARDNESS OF THE STONE COULD NOT BE RULED OUT AS THE CONTRIBUTING FACTOR OF THE USER'S EXPERIENCE. THIS REPORT IS BEING SUBMITTED AS A MEDICAL DEVICE REPORT IN AN ABUNDANCE OF CAUTION.
OLYMPUS WAS INFORMED THAT DURING A THERAPEUTIC ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE, THE SUBJECT DEVICE WAS REPORTEDLY BROKEN AT THE PIPE IN THE PROXIMAL END AS THE LITHOTRIPSY HANDLE WAS CLOSING THE BASKET AROUND A STONE. THE USERS REPORTEDLY ATTEMPTED TO CRUSH THE STONE WITH A SOLENDRA EMERGENCY HANDLE WITHOUT A SUCCESS BUT MANAGED TO REMOVE THE BASKET FROM THE PT. A SIMILAR BUT DIFFERENT DEVICE WAS REPORTEDLY INTRODUCED IN AN ATTEMPT TO CRUSH THE STONE OF WHICH THE DEVICE WAS SAID TO HAVE BROKEN AT THE PIPE IN THE PROXIMAL END AGAIN AND THE SAME SOLENDRA EMERGENCY HANDLE WAS UTILIZED IN WHICH THE PROCEDURE WAS COMPLETED WITH NO PT HARM REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | OLYMPUS SINGLE USE MECHANICAL LITHOTRIPTOR V | LITHOTRIPTOR | LQR | OLYMPUS MEDICAL SYSTEMS CORPORATION | BML-V442QR-30 | 17K |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |